Research of Left Ventricular Assist with Nutating Centrifugal Pump

Research Project

Project/Area Number 02670605
Research Category

Grant-in-Aid for General Scientific Research (C)

Allocation TypeSingle-year Grants
Research Field Thoracic surgery
Research InstitutionKyoto University

Principal Investigator

OKABAYASHI Hitoshi  Kyoto univ. faculty of medicine instructor, 医学部, 助手 (10204009)

Co-Investigator(Kenkyū-buntansha) TAKEUCHI Toshihumi  Kyoto univ. fac. of med. instructor, 医学部, 助手
OKAMOTO Yoshihumi  Kyoto univ. fac. of med. ass prof., 医学部, 助教授 (80116392)
AKAMATSU Teruaki  Kyoto univ. fac. of tech. prof., 工学部, 教授 (40025896)
BAN Toshihiko  Kyoto univ. fac. of med. prof., 医学部, 教授 (00173120)
Project Period (FY) 1990 – 1991
Project Status Completed (Fiscal Year 1991)
Budget Amount *help
¥2,100,000 (Direct Cost: ¥2,100,000)
Fiscal Year 1991: ¥1,200,000 (Direct Cost: ¥1,200,000)
Fiscal Year 1990: ¥900,000 (Direct Cost: ¥900,000)
KeywordsTeaspoon pump / Nutation / Hemolysis / Muscle damage / Tea Spoon Pump / 左心補助 / 血液損傷 / 非拍動流
Research Abstract

The nutating centrifugal pump, dubbed the teaspoon pump(TS pump), was developed by Bauermister and Affeld[1]. Later, it was improved by Akamtsu and colleagues at Kyoto University Faculty of Technology[2-4]. We have recently perfoned experirimental studies in vivo with the TS pump, and examined its safety and efficacy. Nine male sheep, weighing 48-64kg, wree used in this study. Drainage and retum cannulas were inserted into the left atrium and the descending aorta, respectively. The TS pump was then driven in totally assisted circulation in first four experiments, and in partial assist(almost 50%)in the latter experiments. In all cases, the entire circulation was easily maintained, and the mean arterial pressure was maintained at over 65mmHg. Blood samples showed no remarkable abnormality concerning hemolysis ; the values of free hemoglobin 48 hours after the initiation of total assist circulation were 12-13mg/dl. Blood cell counts and total bilirubin values were within normal limits. The antithrombogenicity of the system remained satisfactory. However, some problem were observed : (1)when ventricular fibrillation occurred while the animal was in a hypovolemic state, the drainage was found inadequate ; (2)none of the sheep were able to maintain a standing position postoperatively. In order to achieve more reliable results, we need to resolve these problems.

Report

(3 results)
  • 1991 Annual Research Report   Final Research Report Summary
  • 1990 Annual Research Report
  • Research Products

    (17 results)

All Other

All Publications (17 results)

  • [Publications] 武内 俊史: "Tea Spoon Pumpによる完全左心補助の実験的検討" 人工臓器. 20. 676-679 (1991)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T.Takeuchi: "Experimental study of nutating centrifugal blood pump in vivo" ARTIFICIAL HEART 3. 3. 93-99 (1992)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] 岡林 均: "補助循環(LVAD)施行中の免疫能の変動" 日本脳部外科学会誌. 38. 10-15 (1990)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T.Akamatsu: "Performance of a Compact Nutating Centrifugal Blood Pump (Teaspoon Pump)" Artificial Organs. 14. 30-33 (1990)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T.Akamatsu: "Centrifugal Blood Pump with a magnetically Suspended Impeller" Artificial Organs.

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T. Takeuchi et al.: "Experimental study of total assist for left ventricle with tea spoon pump" Japanese J. of Artificial Organs. 20(3). 676-679 (1991)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T. Takeuchi et al.: "Experimental study of nutating blood pump in vivo" ARTIFICTAL HEART 3. 3. 93-99 (1992)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] H. Okabayashi et al.: "Ohanges in the immune response in postoperative patients while running the ventricular assist device (VAD)" Japanese J. of Thor. and Cardio vasc, Surgery. 38(1). 10-15 (1990)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T, Akamatsu: "Performance of a Compact Nutating Centrifugal Blood Pump (Teaspoon Pump)" Aritificial Organs. 14, Suppl 1. 30-33 (1990)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] T, Akamatsu: "Centrifugal Blood Pump with a Magnetically suspended Impeller" Artificial Organs.

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1991 Final Research Report Summary
  • [Publications] 武内 俊史: "Tea Spoon Pumpによる完全左心補助の実験的検討" 人工臓器. 20. 676-679 (1991)

    • Related Report
      1991 Annual Research Report
  • [Publications] T.Takeuchi: "Experimental study of nutating centrifugal blood pump in vivo" ARTIFICAL HEART 3. 3. 93-99 (1992)

    • Related Report
      1991 Annual Research Report
  • [Publications] 岡林 均: "補助循環(LVAD)施行中の免疫能の変動" 日本胸部外科学会誌. 38. 10-15 (1990)

    • Related Report
      1991 Annual Research Report
  • [Publications] T.Akamatsu: "Performance of a Compact Nutating Centrifugal Blood Pump(Teaspoon Pump)" Artificial Organs. 14. 30-33 (1990)

    • Related Report
      1991 Annual Research Report
  • [Publications] T.Akamatsu: "Centrifugal Blood Pump with a Magnetically Suspended Impeller" Artificial Organs.

    • Related Report
      1991 Annual Research Report
  • [Publications] 武内 俊史: "Experimental Study of Left Ventricular Assist with Tea Spoon Pump in vivo" Artificial Heart 3.

    • Related Report
      1990 Annual Research Report
  • [Publications] 武内 俊史: "Tea Spoon Pumpによる完全左心補助の実験的検討" 人工臓器.

    • Related Report
      1990 Annual Research Report