• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to previous page

Computer Aided Dental Cutting System

Research Project

Project/Area Number 06557105
Research Category

Grant-in-Aid for Scientific Research (A)

Allocation TypeSingle-year Grants
Section試験
Research Field 補綴理工系歯学
Research InstitutionTokyo Medical and Dental University

Principal Investigator

MIYAIRI Hiroo  Tokyo Medical and Dental University, Inst Med Dent Engng, Professor, 医用器材研究所, 教授 (50013892)

Co-Investigator(Kenkyū-buntansha) KIKUCHI Masafumi  Tohoku Univ, Faculty of Dentistry, Assistant, 医用器材研究所, 助手 (50250791)
FUKUDA Hideaki  Tokyo Medical and Dental University, Inst Med Dent Engng, Assistant, 医用器材研究所, 助手 (50014163)
NAGAI Mashiro  Tokyo Medical and Dental University, Inst Med Dent Engng, Assistant, 医用器材研究所, 助手 (10013971)
TAKAKUDA Kazuo  Tokyo Medical and Dental University, Inst Med Dent Engng, Associate Professor, 医用器材研究所, 助教授 (70108223)
Project Period (FY) 1994 – 1996
Project Status Completed (Fiscal Year 1996)
Budget Amount *help
¥10,200,000 (Direct Cost: ¥10,200,000)
Fiscal Year 1996: ¥1,900,000 (Direct Cost: ¥1,900,000)
Fiscal Year 1995: ¥4,600,000 (Direct Cost: ¥4,600,000)
Fiscal Year 1994: ¥3,700,000 (Direct Cost: ¥3,700,000)
KeywordsTooth Cutting / Handpiece / Computer Aided Dentistry / Angle Measurement / Tooth Preparation
Research Abstract

1. The relative angles of a handpiece to an arch are successively measured by attaching the three dimensional position sensors to an arch and a handpiece.
2. Calibrations of the angle measuring system reveals that metals used for prosthodontics and the handpiece do not affect the accuracy of the data.
3. Calibrations over the whole range of the clinically possible areas of usage of handpieces, the maximum error in angle measurement is less than 0.72, which is well below the allowable error.
4. The HMD display device is developed which display the amount of rotations to realize the correct angles of the handpiece to the arch.
5. Parallel preparation experiments are carried out with several dentists. The accuracy of the preparation is considerably increased in every cases.
6. Thus it was successively demonstrate that the usefulness of the system.

Report

(4 results)
  • 1996 Annual Research Report   Final Research Report Summary
  • 1995 Annual Research Report
  • 1994 Annual Research Report

URL: 

Published: 1994-04-01   Modified: 2016-04-21  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI