• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to previous page

A Development of a high sensitivity cephalometric radiography using a photon-counting X-ray sensor.

Research Project

Project/Area Number 07672044
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field 病態科学系歯学(含放射線系歯学)
Research InstitutionOsaka University

Principal Investigator

TOKUOKA Osamu  Osaka University, Faculty of Dentistry, Assistant, 歯学部, 助手 (70155513)

Co-Investigator(Kenkyū-buntansha) HAYAMI Akimune  Osaka University, Faculty of Dentistry, Assist.Professor, 歯学部, 助教授 (30028477)
FUCHIHATA Hajime  Osaka University, Faculty of Dentistry, Professor, 歯学部, 教授 (70028728)
Project Period (FY) 1995 – 1996
Project Status Completed (Fiscal Year 1996)
Budget Amount *help
¥2,200,000 (Direct Cost: ¥2,200,000)
Fiscal Year 1996: ¥300,000 (Direct Cost: ¥300,000)
Fiscal Year 1995: ¥1,900,000 (Direct Cost: ¥1,900,000)
KeywordsDigital Radiography / Cephalometric Radiography / Cd / Te Sensor / デジタルラジオグラフィー / 量子計数型X線センサ / X線画像ワークステーション / 量子計数型X線センサー / CdTe素子 / セファログラム / 骨塩定量 / スキャノグラフィー / ワンショットデュアルエネルギーサブトラクション
Research Abstract

Cd/Te X-ray sensor was used as a super high sensitive X-ray detector for the development of cephalometric radiography. A Cd/Te crystal, a high speed pulse amplifier, acomparator and a 16bit counter were basic components of a X-ray detector. First, a Cd/Te crystal catches a X-ray photon and makes a electric pulse signal wich voltage is in proportion to the X-ray photon energy. Second, a high speed pulse amplifier amplifies that electrical pulse. Third, noise component of electrical signal is cutting off by comparator. Forth, pulse of purified electrical signal is counted by the comparator. In this way, a pixel values are measured. 2 dimensional sensor which size was about 2x130 mm was made. This sensor was equipped on the specially made dental panoramic X-ray machine AZ3000^<TM>. This machine could perform scanography, then the cephalometric radiography was taken as a scanned radiography technique. This system was almost 20 times high sensitive compared with a conventional screen-film system.
To watch a cephalometric images, X-ray image viewing station was made. This viewing system was based on DICOM3.0 standard and had a Hyper Text base user interface.

Report

(3 results)
  • 1996 Annual Research Report   Final Research Report Summary
  • 1995 Annual Research Report

URL: 

Published: 1995-04-01   Modified: 2016-04-21  