Study of measuring system of x-ray spectrum for reduction of patient dose at medical x-ray diagnosis.

Research Project

Project/Area Number 08672583
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field Medical sociology
Research InstitutionOsaka University

Principal Investigator

MATSUMOTO Masao  Osaka University, Faculty of Medicine, associate professor, 医学部, 助教授 (50149944)

Co-Investigator(Kenkyū-buntansha) KANAMORI Hitoshi  Fukui University of Technology, Electrical Engineering, professor, 工学部, 教授 (70027770)
OGATA Yuji  Osaka University, Faculty of Medicine, research associate, 医学部, 助手 (60281127)
Project Period (FY) 1996 – 1997
Project Status Completed (Fiscal Year 1997)
Budget Amount *help
¥2,300,000 (Direct Cost: ¥2,300,000)
Fiscal Year 1997: ¥1,000,000 (Direct Cost: ¥1,000,000)
Fiscal Year 1996: ¥1,300,000 (Direct Cost: ¥1,300,000)
KeywordsCdTe detector / CdZnTe detector / stripping procedure / diagnostic x-ray spectrum / photon spectrum
Research Abstract

(1) Correction of diagnostic x-ray spectra measured with CdTe and CdZnTE detectors
We have attempted to perform the direct measurement of diagnostic x-ray spectra using low-cost CdTe and CdZnTe detectors, which have been developed recently. Since detected spectra did not coincide with the true photon spectra, a correction by stripping procedure was applied. The results obtained with CdTe detector were less exact than those with a Ge detector, but those with CdZnTe detector agreed with it.
We analyzed mammographic x-ray spectra penetrated through 0.03mmMo or 0.025mmRh filter and objects measured using 50-mu m pinhole colimator system at 28kV and 50mAs. The obtainable results with CdZnTe detector agree with mammographic x-ray spectral shape measured using Ge detector published by Bureau of Radiological Health (BRH) in U.S.A..
(2) Measurement of scattered x-ray spectra
By using a conical collimator and a lead beam stopper, primary and scattered x-ray photon spectra through an object and an an … More ti-scatter grid (abbreviated as grid, hereafter) were measured separately, varying the irradiation field area. From these results, the effects of grids for removing scattered x-rays were able to estimate.
(3) Analysis of optimum radiographic factors
We proposed a new method for determining the optimum tube voltage of a chest radiograph using psychophysical analysis. An optimum tube voltage, by which the highest density in the lung-field and the lowest density in the mediastin just covered the optimum density range, was obtained using x-ray photon spectrum and sensitivity spectrum of the screen.
(4) Analysis of image quality (MTF,granularity) of a screen-film system
We have obtained the tube voltage dependence of the modulation transfer function (MTF) of a rate-erath screen-film system. The values of the MTFs for the front and back sides at spatial frequencies of less than 3mm^<-1> decreased with the tube voltage. We showed that to explain this phenomenon, not only K-x rays but coherent and incoherent scattered x-ray photons yielded in the screens had to be taken into consideration.
The value of psychophysical Wiener spectrum at a frequency of 2.0mm^<-1> can be used for the evaluation of radiographic mottle instead of the psychophysical RMS granularity. The time to obtain the psychophysical Wiener spectrum can be reduced more than it to obtain the psychophysical RMS granularity. Less

Report

(3 results)
  • 1997 Annual Research Report   Final Research Report Summary
  • 1996 Annual Research Report
  • Research Products

    (21 results)

All Other

All Publications (21 results)

  • [Publications] Y.Asai: "Evaluation of radiographic mottle using a single frequency value of the psychophysical Wiener spectrum" Journal of Photographic Science. 44・3. 89-92 (1996)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] A.Takigawa: "Effects of anti-scatter grids on X-ray spectrum and radiographic contrast" Journal of Photographic science. 44・3. 93-98 (1996)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] 松本 政雄: "CdTe,CdZnTe検出器を用いて測定した診断用X線スペクトルの補正" 放射線. 22・3. 63-72 (1996)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] H.Arimura: "Tube voltage dependence of modulation transfer function for a rare-earth screen-film system" Journal of The Society of Photographic Science and Technology of Japan. 60・3. 177-181 (1997)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] 浅井 義行: "心理物理的解析から求めた胸部X線写真の最適撮影管電圧" 日本写真学会誌. 60・3. 189-195 (1997)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] M.Matsumoto: "Direct measurement of mammographic X-ray spectra using CdZnTe detector" Medical & Biological Engineering & Computing. 35・S2. 691- (1997)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] A.Takigawa: "Effects of anti-scatter grids on x-ray spectrum and radiographic contrast" Journal of Photographic Science. 44-3. 93-98 (1996)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] M.Matsumoto: "Correction of diagnostic x-ray spectra measured with CdTe and CdZnTe detectors" Ionization Radiation. 22-3. 63-72 (1996)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] Y.Asai: "Evaluation of radiographic mottle using psychophysical analysis" Journal of Medical Imaging and Information Science. 14-2. 83-95 (1997)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] H.Arimura: "Tube voltage dependence of modulation transfer function for a rare-earth screen-film system" Journal of The Society of Photographic Science and Technology of Japan. 60-3. 177-181 (1997)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] Y.Asai: "Optimum tube voltage of chest radiographs obtained by psychophysical analysis" Journal of The Society of Photographic Science and Technology of Japan. 60-3. 189-195 (1997)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] M.Matsumoto: "Direct measurement of mammographic x-ray spectra using CdZnTe detectors" Medical Physics. 24-6. 1026 (1997)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1997 Final Research Report Summary
  • [Publications] 浅井義行: "心理物理的解析を用いたX線写真粒状の評価" 医用画像情報学会雑誌. 14・2. 83-95 (1997)

    • Related Report
      1997 Annual Research Report
  • [Publications] H.Arimura: "Tube voltage dependence of modulation trausfer dunction for a rare-earth screen-filem system" Journal of The Society of Photographic Science and Technology of Japan. 60・3. 177-181 (1997)

    • Related Report
      1997 Annual Research Report
  • [Publications] 浅井義行: "心理物理的解析から求めた胸部X線写真の最適撮影管電圧" 日本写真学会誌. 60・3. 182-188 (1997)

    • Related Report
      1997 Annual Research Report
  • [Publications] 尾崎吉明: "検出能と撮影能率を考慮した軟X線写真の最適撮影条件" 日本写真学会誌. 60・3. 189-195 (1997)

    • Related Report
      1997 Annual Research Report
  • [Publications] M.Matsumoto: "Direct measurement of mammographic X-ray spectra using CdZnTe detector" Medical & Biological Engineering & Computing. 35-S2. 691- (1997)

    • Related Report
      1997 Annual Research Report
  • [Publications] 有村秀孝、松本政雄他6名: "X-ray tube voltage dependence of Wiener spectra of guantum mottle obtained by determining detective guantum efficiencies of a screen" Optical Engineering. 35・4. 926-937 (1996)

    • Related Report
      1996 Annual Research Report
  • [Publications] 浅井義行、松本政雄他3名: "Evaluation of radiographic motlle using a single frequency Valne of psychophysical Wiener spectrum" Journal of Photographic Science. 44・3. 89-92 (1996)

    • Related Report
      1996 Annual Research Report
  • [Publications] 滝川厚、松本政雄、他4名: "Effects of anti-scatter grids on x-ray spectrum and radiographic contrast" Journal of Photographic Science. 44・3. 93-98 (1996)

    • Related Report
      1996 Annual Research Report
  • [Publications] 松本政雄、金森仁志他2名: "CdTe,CdZnTe検出器を用いて測定した診断用X線スペクトルの補正" 放射線. 22・3. 63-72 (1996)

    • Related Report
      1996 Annual Research Report

URL: 

Published: 1996-04-01   Modified: 2016-04-21  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI