Improvement of the SRA distribution in RF capacitive heating of the deep seated tumors
Project/Area Number |
12670891
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Radiation science
|
Research Institution | Kagoshima University |
Principal Investigator |
HIRAKI Yoshiyuki Kagoshima University, Univercity Hospital, Research (90264423)
|
Co-Investigator(Kenkyū-buntansha) |
YUNOKUTI Kazutomo Kagoshima University, Univercity Engineering, Professor (10094187)
NISHIYAMA Kenryu Kagoshima University, University Medicine, Associate Professor (80264422)
KAWANO Kazunori Kagoshima University, Univercity Hospital, Research Associate (50108750)
|
Project Period (FY) |
2000 – 2001
|
Project Status |
Completed (Fiscal Year 2001)
|
Budget Amount *help |
¥2,100,000 (Direct Cost: ¥2,100,000)
Fiscal Year 2001: ¥600,000 (Direct Cost: ¥600,000)
Fiscal Year 2000: ¥1,500,000 (Direct Cost: ¥1,500,000)
|
Keywords | Distribution of the SAR / RF capacitive heating / Oval electrodes / Dharma doll shaped electrodes / deep seated heating / 温熱療法 / 交互加温 / 腔内加温 / 皮下脂肪 / 過熱防止 |
Research Abstract |
RF capacitive hyperthermia has been used for the treatment of various malignan ttumors combined with radiotherapy and chemotherapy. Though a pair of round electrodes of the optimum size have been used by the position and extension of the tumors, round electrodes often hang out of the surfaces of the human body such as arm, thigh, and neck. We contrived some electrodes with oval, triangle and dharma doll in shape to fit a variety of malignant tumors and investigated the peculiarity of heating using agar phantoms. Though hot spot around the edge of electrodes was slightly stronger in oval electrodes than round electrodes, that of dharmal doll shape electrodes was almost equal to that of round electrodes. Capability of deep position by both oval and dharmal doll shape electrodes was superior to that by round electrodes, as well as selectiveness of treatment site. In clinical use of RF capacative hyperthermia, both oval and dharmal doll shape electrodes seemed useful for the treatment of various malignant tumors.
|
Report
(3 results)
Research Products
(4 results)