Analysis and prevention of microbial contamination of nebulizers

Research Project

Project/Area Number 13670382
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field Public health/Health science
Research InstitutionNara Medical University

Principal Investigator

KATSUI Noriaki  Nara Medical University, School of Medicine, Assistant Professor, 医学部, 講師 (70161073)

Project Period (FY) 2001 – 2002
Project Status Completed (Fiscal Year 2002)
Budget Amount *help
¥500,000 (Direct Cost: ¥500,000)
Fiscal Year 2002: ¥500,000 (Direct Cost: ¥500,000)
KeywordsNebulizer / Microbial contamination / Nosocomial infection / Aerosol
Research Abstract

Nebulizers have been recognized as potential agents of nosocomial respiratory infection. In this investigation, the microbial contamination of nebulizers was analyzed and the effective decontamination methods were explored.
1. Out of 20 in-use hospital nebulizers, eighteen (90%) were contaminated and viable microbes in the medication reservoirs of those nebulizers ranged from 10^1〜10^6 CFU/ml. Glucose nonfermentative Gram-negative bacilli were the predominant organisms isolated.
2. According to the statistical analysis of the methods of handling of the nebulizers, the washing of nebulizers with disinfectants and the cold storage of medication solutions afford simple and effective means for the decontamination of the nebulizers.
3. With respect to the components of the nebulizer, the disinfection of the medication chamber, which generates aerosols, was the most important in order to prevent nosocomial infection. On the other hand, even when corrugated tubes (aerosol tubes) and airfilters were contaminated, no viable organisms were detected in aerosols.
4. No viable organisms were detected in every component of the nebulizers after the disinfection with 0.1% Tego 51, ethanol or 0.1% NaOCl for 15 min at room temperature. Drying was also an effective sterilizing method.
5. The jet nebulizer produces heavier contaminated bacterial aerosols than the ultrasonic nebulizer. The contaminant counts of components of the jet nebulizer were 10^2〜10^3 times higher than those of the ultrasonic nebulizer.
6. The medication cup of the ultrasonic nebulizer can be easily damaged by the ultrasonic action and invaded by the contaminants from the ultrasonic chamber. Therefore, the regular examination of the medication cup is necessary.

Report

(3 results)
  • 2002 Annual Research Report   Final Research Report Summary
  • 2001 Annual Research Report
  • Research Products

    (7 results)

All Other

All Publications (7 results)

  • [Publications] 勝井則明: "ネブライザーの微生物汚染に対する効果的管理"日本手術医学会誌. 23. 82-84 (2002)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2002 Final Research Report Summary
  • [Publications] 勝井則明: "微生物汚染の観点からみたネブライザーの蛇管およびエアフィルターの危険性についての評価"日本手術医学会誌. 23. 112-116 (2002)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2002 Final Research Report Summary
  • [Publications] N. Katsui, M. Manabe: "Proper management of nebulizers for prevention of microbial contamination"J. Jpn. Ass. Oper. Med.. 23. 82-84 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2002 Final Research Report Summary
  • [Publications] N. Katsui, M. Manabe: "Microbial hazard evaluation of corrugated tubes and airfilters attached to nebulizers"J. Jpn. Ass. Oper. Med.. 23. 112-116 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2002 Final Research Report Summary
  • [Publications] 勝井則明: "ネブライザーの微生物汚染に対する効果的管理"日本手術医学会誌. 23. 82-84 (2002)

    • Related Report
      2002 Annual Research Report
  • [Publications] 勝井則明: "微生物汚染の観点からみたネブライザーの蛇管およびエアフィルターの危険性についての評価"日本手術医学会誌. 23. 112-116 (2002)

    • Related Report
      2002 Annual Research Report
  • [Publications] 勝井則明: "微生物汚染の観点からみたネブライザーの蛇管およびエアフィルターの危険性についての評価"日本手術医学会誌. 23(印刷中). (2002)

    • Related Report
      2001 Annual Research Report

URL: 

Published: 2002-04-01   Modified: 2016-04-21  