• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to previous page

Quality Assurance Supporting System for Digital X-ray Image Utilizing Filing System of DICOM-3.0

Research Project

Project/Area Number 13671973
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field 病態科学系歯学(含放射線系歯学)
Research InstitutionTokyo Metropolitan University of Health Sciences (TMUHS)

Principal Investigator

KATOH Tsuguhisa  TMUHS, Faculty of Health Sciences, Associate Professor, 保健科学部, 助教授 (70095107)

Co-Investigator(Kenkyū-buntansha) NEGISHI Tohru  TMUHS, Faculty of Health Sciences, Assistant, 保健科学部, 助手 (00259144)
OHTANI Hiroki  TMUHS, Faculty of Health Sciences, Associate Professor, 保健科学部, 講師 (10259145)
SAITOH Hidetoshi  TMUHS, Faculty of Health Sciences, Associate Professor, 保健科学部, 助教授 (50196002)
SASAKI Takehito  Tokyo Medical and Dental University Graduate School, Emeritus Professor, 大学院・口腔放射線診断部門, 名誉教授 (90013896)
MYOJOYAMA Atsushi  TMUHS, Faculty of Health Sciences, Assistant, 保健科学部, 助手 (90347279)
Project Period (FY) 2001 – 2003
Project Status Completed (Fiscal Year 2003)
Budget Amount *help
¥3,200,000 (Direct Cost: ¥3,200,000)
Fiscal Year 2003: ¥500,000 (Direct Cost: ¥500,000)
Fiscal Year 2002: ¥500,000 (Direct Cost: ¥500,000)
Fiscal Year 2001: ¥2,200,000 (Direct Cost: ¥2,200,000)
KeywordsDigital X-ray Image / Quality Assurance / Radiation Dosimetry / Imaging Plate / Output X-ray Intensity / Half Value Layer / Tube Potential / Film Badge Case / X線の線質 / 品質管理支援ソフトウェア / DICOM規格 / X線量 / ディジタルX線撮影
Research Abstract

A quality control (QC) system for the quality assurance (QA) of digital x-ray image was developed. The system would support the QC procedures with the least labor of staff and cost. The main outcomes of the development were as follows.
1)Measurement of output x-ray intensity, half value layer (HVL) and tube potential utilizing computed radiography(CR) system
An imaging plate (IP) for intraoral radiography was inserted into a film badge case for personal dosimeter with filters made of aluminum, copper and tin on the measuring windows. This system could measure the output x-ray intensity with an error of ±5%, the HVL with an error of ±0.17mm and the tube potential with an error of ±2%.
2)Method for evaluation of IP's degradation
A method for the detection of the scratch noise of IP for intraoral radiography was developed. An IP was uniformly radiated by x-rays and scanned. A digital image processing software was developed to get a binary image of the scratch noise.
3)Development of QA supporting information system
A QA supporting information system based on DICOM standard was developed. Based on the object oriented analysis of devices and materials concerning QC, the structure of the information objects and performance of the servers and clients were defined. As a result, the specification was applicable not only to dental radiography but also to all the modality used in radiology.

Report

(4 results)
  • 2003 Annual Research Report   Final Research Report Summary
  • 2002 Annual Research Report
  • 2001 Annual Research Report
  • Research Products

    (19 results)

All Other

All Publications (19 results)

  • [Publications] 加藤二久, 西谷源展, 他: "診断X線用線量計の校正センター発足にあたって"医用線量標準. 8. 47-51 (2002)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, T.Sasaki, et al.: "A Method for Evaluation of Entrance Surface Dose from the Measurement of Exposure and HVL in Intraoral radiography"Radiat.Protec.Dosim.. 103. 47-55 (2003)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, H.Saitoh, et al.: "Computer Aided Quality Assurance for Medical Imaging : Proposal of DICOM QA"World Cong.Med.Phys.Biomed.Engin.2003 Sydney. TH TR6 168 (2003)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, H.Saitoh, et al.: "A Method for Simultaneous measurement of Air kerma, HVL and Tube Potential in Quality Control Procedure"Proc.3rd KJMP, 2nd Asia Oceania Cong.Med.Phys.. 294-297 (2002)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, K.Imamura, T.Matsumoto: "Current Topics on Quality Assurance of X-ray Diagnosis"Proc.3rd KJMP, 2nd Asia Oceania Cong.Med.Phys.. 13-16 (2002)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] 加藤二久, 加藤洋, 他: "東京都立保健科学大学における診断X線用線量計校正の再現性"医学物理. 22 Suppl.1. 260-262 (2002)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, M.Mishitani, S.Koyama, M.Kumagai, S.Iwanami, T.Kikuchi, Y.Katoh, H.Maekoshi: "On the Calibration Center for Diagnostic X-ray Dosimeters (in Japanese)"Medical Dose Standards. 8(1). 47-54 (2003)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, T.Sasaki, K.Iwai, T.Okano, K.Satoh, T.Shimano, A.Hayami, S.Wada, N.Juto: "A Method for Evaluation of Entrance Surface Dose.from the Measurement of Exposure and HVL in Intraoral Radiography using Radiophotoluminecent Dosimeter."Radiat.Protec.Dosim.. 103(1). 47-55 (2003)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, H.Saitoh, T.Negishi, H.Ohtani, A.Myojoyama, Y.Ohno, Y.Kimura, A.Mori, T.Sasaki: "Computer Aided Quality Assurance for Radiological Imaging : Proposal of DICOM QA."World Cong.Med.Phys.Biomed.Engin.. Sydney TH TR6 168. (2003)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, H.Saitoh, H.Ohtani, T.Negishi, A.Myojoyama, Y.Ohno, T.Sasaki: "A method for simultaneous measurement of air kerma, half value layer and tube potential in quality control procedure of diagnostic x ray units."Proc.3rd KJMP and 2nd Asia Oceania Cong.Med.Phys.. 294-297 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, K.Imamura, T.Matumoto: "Current Topics on Quality Assurance of X-ray Diagnosis in Japan."Proc.3rd KJMP and 2nd Asia Oceania Cong.Med.Phys.. 13-16 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      2003 Final Research Report Summary
  • [Publications] T.Katoh, H.Saitoh, T.Negishi, H.Ohtani, A.Myojoyama, Y.Ohno, Y.Kimura, et al.: "Computer Aided Quality Assurance for Radiological Imaging : Proposal of DICOM QA"Proc.World Cong.Med.Phys.Biol.Med.Engin.2003 Sydney. THTR6-THTR168 (2003)

    • Related Report
      2003 Annual Research Report
  • [Publications] 木村由美, 加藤二久, 遠藤敦, 三島章, 森田康彦, 岡野友宏 他: "歯科用ディジタルX線画像の画質管理第一報イメージングプレートの劣化について"歯科放射線. 14 Suppl.. 73 (2003)

    • Related Report
      2003 Annual Research Report
  • [Publications] 加藤二久, 木村由美, 遠藤敦, 小林馨, 森田康彦, 岡野友宏 他: "DICOMを利用した放射線画像の質的保証支援プログラム"歯科放射線. 14 Suppl.. 74 (2003)

    • Related Report
      2003 Annual Research Report
  • [Publications] T.Katoh, H.Saitoh, et al.: "A method for simultaneous measurement of air Kerma halfvalue layer and tube potential in quality control procedur of diagnostic"Proc. 2nd Asia Oceania Cong. Med. Phys.. 294-297 (2002)

    • Related Report
      2002 Annual Research Report
  • [Publications] T.Katoh, K.Iwamura, T.Matumoto: "Qurrent Topics of X ray Diagnosis in Japan"Proc. 2nd Asia Oceania Cong. Med. Phys.. 13-16 (2002)

    • Related Report
      2002 Annual Research Report
  • [Publications] T.Katoh, T.Sasaki, et al.: "A method for evaluation of entrance surface dose from the measurement of exposure and HVL in intra-oral radiography using radiophotoluminescent dosemeter"Radiat. Protec. Dosim.. 103. 47-55 (2003)

    • Related Report
      2002 Annual Research Report
  • [Publications] 加藤二久, 根岸徹, 大谷浩樹 他: "東京都立保健科学大学における診断用線量計較正の再現性"医学物理. 22(Supl.2)(発行予定). (2002)

    • Related Report
      2001 Annual Research Report
  • [Publications] 加藤二久, 西谷源展, 小山修司 他: "診断領域X腺様線量計のトレーサブルシステム構築にむけて"日本放射線技術学会誌. 57(12). 1438-1443 (2001)

    • Related Report
      2001 Annual Research Report

URL: 

Published: 2001-04-01   Modified: 2016-04-21  