Budget Amount *help |
¥4,680,000 (Direct Cost: ¥3,600,000、Indirect Cost: ¥1,080,000)
Fiscal Year 2017: ¥910,000 (Direct Cost: ¥700,000、Indirect Cost: ¥210,000)
Fiscal Year 2016: ¥2,340,000 (Direct Cost: ¥1,800,000、Indirect Cost: ¥540,000)
Fiscal Year 2015: ¥1,430,000 (Direct Cost: ¥1,100,000、Indirect Cost: ¥330,000)
|
Outline of Final Research Achievements |
PET detectors have high high performance of sensitivity and spatial resolution. PET detectors can be further improved by tracking the interaction of gamma ray by using Compton camera. In this work, we developed the new Compton tracking PET detector. This detector consist of the scatterer part of the GAGG array and the MPPC array and the absorber part of the GSO array and the PS-PMT. For scatterer part, 0.9 mm scintillator can be identified easily and energy resolution was 15%. The developed detector promises high performance for single gamma based medical imaging.
|