Development of MRI-compatible brachytherapy applicator for elderly asian gynecologic cancer paatients
Project/Area Number |
15K19792
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Radiation science
|
Research Institution | Kobe University |
Principal Investigator |
Uezono Haruka 神戸大学, 医学研究科, 助教 (60546480)
|
Project Period (FY) |
2015-04-01 – 2017-03-31
|
Project Status |
Completed (Fiscal Year 2016)
|
Budget Amount *help |
¥3,900,000 (Direct Cost: ¥3,000,000、Indirect Cost: ¥900,000)
Fiscal Year 2016: ¥2,340,000 (Direct Cost: ¥1,800,000、Indirect Cost: ¥540,000)
Fiscal Year 2015: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
|
Keywords | 子宮頸癌 / 画像誘導小線源治療 / アプリケーター / 腔内照射 / 子宮頸がん |
Outline of Final Research Achievements |
We conducted a study to verify the safety of the prototype applicator. On physical experiment, we confirmed the radiation effect to the surrounding tissue was equal to that of commercially available applicator. However, extremely high dose area was found to happen probably due to the short distance between each needle. The applicator will be connected to remote after loading system (RALS) in clinical practice. Hence the connection needs to be certain so that the radioactive seed can be carried through the unit without any interruption. We have found the connection was smooth. To make it commercially available, collaboration with RALS trader is essential. We are going to build up a collaborative relationship with the company, which has not yet been carried out at the moment.
|