Outline of Final Research Achievements |
Generally, the artificial heart refers to the left ventricular assist device, designed to assist the left ventricle. In this study, however, we worked on the development of an artificial heart that can support or replace both ventricular with a single rotor. In this research, we mainly researched and developed a magnetic levitation pump. We clarified that the manufactured magnetic levitation pump achieved a pumping capacity of 100 mmHg and a flow rate of 5.8 L/min at a nominal rotation speed of 1800 r/min. We clarified the pump could support both ventricles.
|