Establishment of new fracture resistance testing and non-destructive measurement methods using micro-CT and new measuring devices
Project/Area Number |
20K18618
|
Research Category |
Grant-in-Aid for Early-Career Scientists
|
Allocation Type | Multi-year Fund |
Review Section |
Basic Section 57050:Prosthodontics-related
|
Research Institution | Nihon University |
Principal Investigator |
|
Project Period (FY) |
2020-04-01 – 2022-03-31
|
Project Status |
Completed (Fiscal Year 2021)
|
Budget Amount *help |
¥2,470,000 (Direct Cost: ¥1,900,000、Indirect Cost: ¥570,000)
Fiscal Year 2021: ¥1,170,000 (Direct Cost: ¥900,000、Indirect Cost: ¥270,000)
Fiscal Year 2020: ¥1,300,000 (Direct Cost: ¥1,000,000、Indirect Cost: ¥300,000)
|
Keywords | マイクロCT / ジルコニア / インプラント上部構造 / 破壊強度試験 / 間接修復用コンポジットレジン / 非破壊的測定法 / 破壊強度 |
Outline of Research at the Start |
口腔インプラント治療において、ジルコニアが用いられる機会が多くなってきたが、前装陶材の微小破折が報告されており、それに起因する咬合接触の変化や下顎位の変位によりインプラント治療の予後に大きく影響することが懸念される。しかし、従来の破壊強度試験では、試験後に荷重が解放されてしまい、破壊に至る亀裂の起点や進展を経時的に観察することはできない。そこで、補綴装置の内面を非破壊的に撮影および観察が可能なin vivo micro-CT装置と荷重を負荷したまま破壊強度試験が可能な新規試験用測定器を用い、破壊に至る亀裂の起点と進展を経時的に観察する試験法を考案した。
|
Outline of Final Research Achievements |
A new measuring device for the test was designed with a threading on the inner surface of the measuring device to allow micro-CT imaging without releasing the load by rotating the load part and applying stress. In addition, the applied load was recorded by a load cell. The results of the measurements using the new test device showed that the ceramic-prepared zirconia prosthetic device had reached fracture with no deformation or other significant changes in the superstructure. On the other hand, in the case of the composite resin-applied zirconia prosthetic device for indirect restorations, the mechanism leading to fracture was observed to differ depending on the material used, such as the development of cracks in the prosthetic material before fracture and deformation of the prosthetic material due to stress.
|
Academic Significance and Societal Importance of the Research Achievements |
本研究で製作した新規試験用測定器を用いた測定法によって、従来の万能試験機を用いた破壊強度試験では観察することができなかった、補綴装置内部に生じるクラックや前装材料の変形が観察された。さらに、micro-CTを応用した測定法より、非破壊的に補綴装置の内面性状を観察することが可能となった。これは、今までの基礎的研究に新たな知見を加えるものである。 本研究で評価された補綴装置は臨床応用可能な破壊強度を有しており、内面性状を観察できる測定方法と合わせて評価することで、安定した補綴装置の設計および製作が可能となり、良好な治療結果および口腔環境の維持に役立つことが期待される。
|
Report
(3 results)
Research Products