Budget Amount *help |
¥18,590,000 (Direct Cost: ¥14,300,000、Indirect Cost: ¥4,290,000)
Fiscal Year 2011: ¥3,770,000 (Direct Cost: ¥2,900,000、Indirect Cost: ¥870,000)
Fiscal Year 2010: ¥8,710,000 (Direct Cost: ¥6,700,000、Indirect Cost: ¥2,010,000)
Fiscal Year 2009: ¥6,110,000 (Direct Cost: ¥4,700,000、Indirect Cost: ¥1,410,000)
|
Research Abstract |
With an electron microscope nano-tomographic images(nano-order resolution) of various materials can be taken, in which the same principle as medical X-ray CT apparatuses is used. However, in general, because a specimen is put on a support thin film extended horizontally the tilt angle is limited to about 60deg.(90deg. ideal). Therefore, only distorted tomographic images are obtained. In this research work, a novel image reconstruction method was devised. This was applied to them and as a result successful images unaffected by the distortion artifact were obtained. Furthermore, it was achieved that the electron dose was reduced to about 20% as compared with the general case.
|