Development of Neonatal peripheral circulation monitor
Project/Area Number |
22591206
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Embryonic/Neonatal medicine
|
Research Institution | Wakayama Medical University |
Principal Investigator |
|
Co-Investigator(Kenkyū-buntansha) |
HIGUCHI Ryuzo 和歌山県立医科大学, 医学部, 准教授 (70156579)
YOSHIKAWA Norishige 和歌山県立医科大学, 医学部, 教授 (10158412)
|
Project Period (FY) |
2010-04-01 – 2013-03-31
|
Project Status |
Completed (Fiscal Year 2013)
|
Budget Amount *help |
¥4,810,000 (Direct Cost: ¥3,700,000、Indirect Cost: ¥1,110,000)
Fiscal Year 2012: ¥650,000 (Direct Cost: ¥500,000、Indirect Cost: ¥150,000)
Fiscal Year 2011: ¥650,000 (Direct Cost: ¥500,000、Indirect Cost: ¥150,000)
Fiscal Year 2010: ¥3,510,000 (Direct Cost: ¥2,700,000、Indirect Cost: ¥810,000)
|
Keywords | 末梢循環 / レーザードップラー血流計 |
Research Abstract |
Methods: Skin blood flow waveforms of neonates were acquired with laser doppler flowmeter. Maximum slope value of each waveform (Smax) was calculated. Results: In control group with stable vital signs, mean Smax was 1.71 (V/s). In a premature baby (gestational age: 25 weeks, birth weight: 927 g) who suffered patent ductus arteriosus and underwent clipping operation on day 25, we acquired 4 time-series samples. The Smax values were 2.0 (day 17), 3.7 (day 20), 4.6 (day 24; the day before operation), 1.4 (day 26; the next day of the operation). Conclusion: We visualized the vascular resistance of capillary vessels with laser doppler flowmeter.
|
Report
(4 results)
Research Products