Project/Area Number |
23500529
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Biomedical engineering/Biological material science
|
Research Institution | Kitasato University |
Principal Investigator |
|
Project Period (FY) |
2011 – 2013
|
Project Status |
Completed (Fiscal Year 2013)
|
Budget Amount *help |
¥5,200,000 (Direct Cost: ¥4,000,000、Indirect Cost: ¥1,200,000)
Fiscal Year 2013: ¥520,000 (Direct Cost: ¥400,000、Indirect Cost: ¥120,000)
Fiscal Year 2012: ¥1,040,000 (Direct Cost: ¥800,000、Indirect Cost: ¥240,000)
Fiscal Year 2011: ¥3,640,000 (Direct Cost: ¥2,800,000、Indirect Cost: ¥840,000)
|
Keywords | OCT / 医用生体工学 / 光干渉断層計測 / 偏光感受型OCT / 生体計測 / 前眼部OCT / 生体医工学 / 偏光計測 / 偏光 |
Research Abstract |
We developed a high-speed, high-sensitivity, swept-source optical coherence tomography (SS-OCT) system with a depth range of 12 mm and a fast A-scan rate of 20,000 times per second, providing high-quality imaging of the whole anterior segment. Using this SS-OCT system, we measured curvature of the anterior and posterior surfaces of the lens, thickness of the lens, and anterior chamber depth in response to far (0.4 D) and near (10 D) accommodative stimulus targets. The radius of curvature of the anterior and posterior surface of the human lens became smaller with accommodation, and changes in the anterior surface were larger than the posterior. Morphometric changes of the lens with accommodation reduced with aging, except for posterior surface curvature.
|