Development of Signal Recovery Method in Space Charge Measurement
Project/Area Number |
23560378
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Electronic materials/Electric materials
|
Research Institution | Matsue National College of Technology |
Principal Investigator |
FUKUMA Masumi 松江工業高等専門学校, 電気工学科, 教授 (70228930)
|
Project Period (FY) |
2011 – 2013
|
Project Status |
Completed (Fiscal Year 2013)
|
Budget Amount *help |
¥5,070,000 (Direct Cost: ¥3,900,000、Indirect Cost: ¥1,170,000)
Fiscal Year 2013: ¥1,820,000 (Direct Cost: ¥1,400,000、Indirect Cost: ¥420,000)
Fiscal Year 2012: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
Fiscal Year 2011: ¥1,690,000 (Direct Cost: ¥1,300,000、Indirect Cost: ¥390,000)
|
Keywords | 空間電荷 / 高電圧 / 誘電体材料 / 音波 / 計測 / 音響インピーダンス / パルス静電応力法 / 減衰 / 誘電体 / PEA法 / 信号補正 / 補正 |
Research Abstract |
Pulsed electro-acoustic (PEA) method is widely used for space charge distribution in insulating materials. The simultaneous measurements of space charge signal and acoustic properties signal within a sample in PEA system have been assembled by adding an Acoustic-impedance microscope (AIM). By using the experimentally assembled calibration system, a calibration method for the PEA method was used successfully to recover the acoustic signal in space charge measurement. The method uses the relationship between the PEA signal and the AIM signal. The calibration system can solve problems such as attenuation, acoustic impedance mismatching between upper electrode and sample. Both a semi-conductor and brass were used as the upper electrode and both materials gave the same results. The calibration method was therefore found to be an effective means of calibrating the measured PEA signal. Furthermore this calibration method doesn't have to use the semi-conductor.
|
Report
(4 results)
Research Products
(28 results)