The aim of this study is to evaluate a X-ray microbeam dosimetry at the synchrotron radiation facilities SPring-8. We have focused on the dose measurements in inhomogeneous phantom geometries. The microscope with CCD camera was used to measure the optical-density distribution of GafChromic films irradiated with multi-slit microbeam X-rays. The peak and valley dose rates in inhomogeneous phantom geometries were found to differ according to density variation in phantom. This result suggests that the microbeam dosimetry is required with evaluation of biological effects in microbeam radiation therapy.