Budget Amount *help |
¥5,070,000 (Direct Cost: ¥3,900,000、Indirect Cost: ¥1,170,000)
Fiscal Year 2014: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
Fiscal Year 2013: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
Fiscal Year 2012: ¥1,950,000 (Direct Cost: ¥1,500,000、Indirect Cost: ¥450,000)
|
Outline of Final Research Achievements |
We designed and produce an equipment for an ureterolith test using a real or model ureter stent and estimated effect of small pores geometry of the stent on ureterolith formation in comparison with CFD analysis of flow in the stents. As a result, it was found that somewhat difference was confirmed in ureterolith formation though a flow field in the stent was independent on the pore shapes in such low Reynold’s number condition. We also estimated behavior of a laser induced bubble for a laser lithotripsy. A bubble induced by a fiber-laser near a solid wall was observed by a high speed video camera under various laser irradiation conditions. As a result, characteristic bubble collapse behaviors and impacts toward a solid wall were obtained.
|