Comparison of Liquid Crystal Display (LCD) of Medical-grade and General-grade LCD, and Tablet PC: Observer Performance Study
Project/Area Number |
24591753
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Radiation science
|
Research Institution | Nagoya University |
Principal Investigator |
YOSHIMURA Kumiko 名古屋大学, 医学(系)研究科(研究院), 教務職員 (90419151)
|
Co-Investigator(Kenkyū-buntansha) |
NIHASHI Takashi 名古屋大学, 医学系研究科, 講師 (00426496)
|
Project Period (FY) |
2012-04-01 – 2015-03-31
|
Project Status |
Completed (Fiscal Year 2014)
|
Budget Amount *help |
¥5,330,000 (Direct Cost: ¥4,100,000、Indirect Cost: ¥1,230,000)
Fiscal Year 2014: ¥1,040,000 (Direct Cost: ¥800,000、Indirect Cost: ¥240,000)
Fiscal Year 2013: ¥1,170,000 (Direct Cost: ¥900,000、Indirect Cost: ¥270,000)
Fiscal Year 2012: ¥3,120,000 (Direct Cost: ¥2,400,000、Indirect Cost: ¥720,000)
|
Keywords | 医用モニタ / 読影実験 / 輝度 / 解像度 / 液晶モニタ / タブレット端末 / 医用画像 / 画像診断 / タブレット / モニタ診断 |
Outline of Final Research Achievements |
The purpose of this study was to compare the reliability in medical use of radiographic image interpretation using medical-grade liquid crystal displays (LCDs) and general-grade LCDs (including the Tablet personal computer (PC)) by examining the observers’ performance and physical performance evaluation. When evaluating and comparing the performance of monitors, adopting luminance, contrast, modulation transfer function (MTF), and noise power spectrum (NPS) between medical-grade LCD and tablet PCs, the tablet PCs performance was equivalent to that of the medical-grade LCD. Observer performance for detecting cerebral infarction of brain computed tomography (CT) was higher on the medical-grade LCD than Tablet PC. Therefore, the Tablet PC could not substitute for medical-grade LCD monitor. However, as tablet PC has promising potentials such as portability, further examination is needed for the clinical situation using tablet PC.
|
Report
(4 results)
Research Products
(4 results)
-
[Journal Article] Comparison of liquid crystal display monitors calibrated with gray-scale standard display function and with γ 2.2 and iPad: observer performance in detection of cerebral infarction on brain CT.2013
Author(s)
Yoshimura K, Nihashi T, Ikeda M, Ando Y, Kawai H, Kawakami K, Kimura R, Okada Y, Okochi Y, Ota N, Tsuchiya K, Naganawa S.
-
Journal Title
American Journal of Roentgenology
Volume: 200(6)
Issue: 6
Pages: 1304-9
DOI
Related Report
Peer Reviewed
-
[Journal Article] Comparison of Liquid Crystal Display (LCD) Calibrated with the Grayscale Standard Display Function (GSDF) and with Gamma 2.2 and iPad: Observer Performance in Detection of Cerebral Infarction on Brain CT.2013
Author(s)
Kumiko Yoshimura, Takashi Nihashi, Mitsuru Ikeda Yoshio Ando, Hisashi Kawai, Kenichi Kawakami, Reiko Kimura, Yumiko Okada, Yoshiyuki Okochi, Naotoshi Ota, Kenichi Tsuchiya, Shinji Naganawa
-
Journal Title
American Journal of Roentgenology
Volume: in press
Related Report
Peer Reviewed
-
-