Establishment of accuracy management system in IMRT QA/QC using organic phosphor
Project/Area Number |
24791336
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Radiation science
|
Research Institution | Keio University |
Principal Investigator |
|
Project Period (FY) |
2012-04-01 – 2014-03-31
|
Project Status |
Completed (Fiscal Year 2013)
|
Budget Amount *help |
¥2,470,000 (Direct Cost: ¥1,900,000、Indirect Cost: ¥570,000)
Fiscal Year 2013: ¥1,040,000 (Direct Cost: ¥800,000、Indirect Cost: ¥240,000)
Fiscal Year 2012: ¥1,430,000 (Direct Cost: ¥1,100,000、Indirect Cost: ¥330,000)
|
Keywords | 強度変調放射線治療 / 品質保証 / 品質管理 / 品質保証・管理 / 有機蛍光体 |
Research Abstract |
In this study, the dose and mechanical accuracy in irradiation field for intensity modulated radiation therapy using the charge coupled device camera-organic phosphor with area type detector system was evaluated. The area type detector, which is made from plastic scintillator known as water equivalent material, and reflecting mirror were fabricated. Acquisitions of information such as stopping positions and moving distance related to multileaf collimator, and absorbed dose were able to obtain simultaneously with high precision easily. The system was achieved within the precision at clinical adaptive level, and the improvement of the quality assurance and more safely treatment were expected.
|