Performance evaluation of lung cancer CT screening CAD and developmant
Project/Area Number |
25461803
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Radiation science
|
Research Institution | Niigata University |
Principal Investigator |
|
Co-Investigator(Kenkyū-buntansha) |
OHKUBO Masaki 新潟大学, 医歯学系, 教授 (10203738)
|
Research Collaborator |
MATSUMOTO Toru
MURAO Kohei
MARASINGHE Janaka C
KOBAYASHI Hajime
YOSHIDA Akifumi
GOUDO Makoto
NARITA Akihiro
SUGAWARA Hideyoshi
|
Project Period (FY) |
2013-04-01 – 2016-03-31
|
Project Status |
Completed (Fiscal Year 2015)
|
Budget Amount *help |
¥4,940,000 (Direct Cost: ¥3,800,000、Indirect Cost: ¥1,140,000)
Fiscal Year 2015: ¥1,300,000 (Direct Cost: ¥1,000,000、Indirect Cost: ¥300,000)
Fiscal Year 2014: ¥1,300,000 (Direct Cost: ¥1,000,000、Indirect Cost: ¥300,000)
Fiscal Year 2013: ¥2,340,000 (Direct Cost: ¥1,800,000、Indirect Cost: ¥540,000)
|
Keywords | CAD / Quality Assurance / Point Spread Function / CT(Computed Tomography) / Virtual nodule / Lung cancer CT screening / 3D-PSF / Low dose CT screening / point spread function / quality assurance of CAD / virtual nodule / lung cancer CT screening / Low Dose CT Screening / Lung Cancer CT Screening / Quality Assurance of CAD / 3D-PSF / Virtual Nodule / ROC / CT screening / PSF / CAD performance / lung cancer |
Outline of Final Research Achievements |
AAPM-CADSC have published the CAD- Quality Assurance (QA) guideline in Medical Physics recently. We also believe the study of CAD-QA is essential to establish the high quality lung cancer CT screening. The purpose of this study is to develop a method of QA of lung cancer CT screening CAD systems. The research strategy is aimed to cover the image quality characteristic problem caused by a variety of CT image quality due to the difference of the CT manufactures, image acquisition and reconstruction parameters. PSF and SSP were measured to know the 3D-PSF. Using 3D-PSF the virtual nodules were generated. The following 4 issues were discussed in applying the virtual nodule for CAD-QA, (1) Verification of PSF-based virtual nodule method using a CT screening database, (2) virtual nodule application to CAD study as a performance evaluation method, (3) study for the sophistication of the Virtual Nodule,(4) robustness evaluation of CAD performance of algorithm differences.
|
Report
(4 results)
Research Products
(26 results)
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
[Presentation] 胸部CT検診CAD性能評価におけるnodule offsetの影響2014
Author(s)
Akifumi Yoshida,Masaki Ohkubo,Hajime Kobayashi,Marasinghe Janaka C,D.M.Janaka Nuwan,Kohei Murao,Toru Matsumoto,Shunsuke Sone, and Shinihi Wada
Organizer
第107会日本医学物理学会学術大会
Place of Presentation
パシフィコ横浜
Year and Date
2014-04-10 – 2014-04-13
Related Report
-
-
-
-
-
-