Coronary FFR Estimation method using Dynamic Coronary CT images
Project/Area Number |
25861107
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Radiation science
|
Research Institution | Hiroshima University |
Principal Investigator |
Toru Higaki 広島大学, 医歯薬保健学研究院(医), 特任助教 (80611334)
|
Project Period (FY) |
2013-04-01 – 2016-03-31
|
Project Status |
Completed (Fiscal Year 2015)
|
Budget Amount *help |
¥4,160,000 (Direct Cost: ¥3,200,000、Indirect Cost: ¥960,000)
Fiscal Year 2015: ¥1,040,000 (Direct Cost: ¥800,000、Indirect Cost: ¥240,000)
Fiscal Year 2014: ¥1,300,000 (Direct Cost: ¥1,000,000、Indirect Cost: ¥300,000)
Fiscal Year 2013: ¥1,820,000 (Direct Cost: ¥1,400,000、Indirect Cost: ¥420,000)
|
Keywords | 冠動脈CT / CT-FFT / ダイナミックCT / CT-FFR / 画像位置あわせ / 3Dプリンタ / 冠動脈血流予備量比 / ダイナミック心臓CT / 心臓CT / FFR |
Outline of Final Research Achievements |
In this study, we developed a method to estimate Coronary Flow Reserve (FFR), which is one of index of cardiac function using dynamic myocardial CT images. We used images acquired from myocardial perfusion study. We estimated the FFR value from shape and time-dependent change of the CT number on coronary region. We evaluated accuracy of our developed method using coronary model phantom made from 3D printing system. In results of the evaluation, we confirmed that the FFR value was dropped after simulated stenosis region. In addition, we found that the FFR value is dropped after normal branching.
|
Report
(4 results)
Research Products
(1 results)