Anthropomorphic phantom based on patient CT
Project/Area Number |
25861117
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Radiation science
|
Research Institution | Nagoya City University |
Principal Investigator |
Omachi Chihiro 名古屋市立大学, 医学(系)研究科(研究院), 研究員 (20588967)
|
Project Period (FY) |
2013-04-01 – 2016-03-31
|
Project Status |
Completed (Fiscal Year 2015)
|
Budget Amount *help |
¥4,290,000 (Direct Cost: ¥3,300,000、Indirect Cost: ¥990,000)
Fiscal Year 2015: ¥390,000 (Direct Cost: ¥300,000、Indirect Cost: ¥90,000)
Fiscal Year 2014: ¥910,000 (Direct Cost: ¥700,000、Indirect Cost: ¥210,000)
Fiscal Year 2013: ¥2,990,000 (Direct Cost: ¥2,300,000、Indirect Cost: ¥690,000)
|
Keywords | 人体等価ファントム / 粒子線治療 / 3Dプリンタ / 3Dプリンタ |
Outline of Final Research Achievements |
The purpose of this study is establishing the verification method of dose distribution in the human body using the anthropomorphic phantom based on patient CT. The 3D printer was used making patient specific anthropomorphic phantom to reproduce the tumor and body shape. The reproducibility of patient structure was verified by measuring the dose distribution. The patient specific anthropomorphic phantom consists of two materials, a plaster bone which made by 3D printer and gelatin for the water equivalent tissue. The measured dose distribution at the air region in the body did not match the patient plan. But the gamma analysis with 3 mm/3 % was 95.4 % in an irradiation field. From these results, the method of this study could reproduce the body structures to the anthropomorphic phantom.
|