Budget Amount *help |
¥4,160,000 (Direct Cost: ¥3,200,000、Indirect Cost: ¥960,000)
Fiscal Year 2016: ¥650,000 (Direct Cost: ¥500,000、Indirect Cost: ¥150,000)
Fiscal Year 2015: ¥650,000 (Direct Cost: ¥500,000、Indirect Cost: ¥150,000)
Fiscal Year 2014: ¥2,860,000 (Direct Cost: ¥2,200,000、Indirect Cost: ¥660,000)
|
Outline of Final Research Achievements |
We developed 3D imaging system by combining micro-PIXE and a X-ray detector with a wave guide. Because only the X-rays which are generated from the plane that is parallel to a wave guide are detected by the system, the beam position has the information of the depth. Therefore three dimensional elemental distribution could be obtained by combining the sample movement. The depth resolution is estimated around 60um. Since the X-rays from deep position of the sample are absorbed, this method could not apply thick sample. This method will be effective for the analysis of the sample that has gaps and holes.
|