Development of the evaluation method for oral function using 320-ADCT
Project/Area Number |
26870687
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Medical Physics and Radiological Technology
Social dentistry
|
Research Institution | Fujita Health University |
Principal Investigator |
|
Project Period (FY) |
2014-04-01 – 2016-03-31
|
Project Status |
Completed (Fiscal Year 2015)
|
Budget Amount *help |
¥3,120,000 (Direct Cost: ¥2,400,000、Indirect Cost: ¥720,000)
Fiscal Year 2015: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
Fiscal Year 2014: ¥1,560,000 (Direct Cost: ¥1,200,000、Indirect Cost: ¥360,000)
|
Keywords | 医用画像情報学 / 口腔機能 / CT被ばく / 嚥下機能 / リハビリテーション医学 / 放射線 / 医療 / 生体機能利用 / リハビリ / ADCT / 嚥下 / 被ばく |
Outline of Final Research Achievements |
The purpose of this study was to investigate the evaluation method for oral function using 320-ADCT. We investigated the effect of the latest dose reduction system. We used a tube voltage of 120 kV and a tube current of 40 mA and performed a dynamic volume scan for a period of 3.3 s. The result for the effective dose was 1.08mSv. The oral function CT examination with double contrast method conditions is tube voltage of 120 kV and a tube current of 10 mA and performed a dynamic volume scan for a period of 9.9 s. The result for the effective dose was 1.5mSv. The results obtained in this study would become the useful information when we reconsidered the radioprotection and the optimization of oral function CT examinations.
|