• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to previous page

Full-automatic measurement system for the corneal endothelial cell morphology in vivo

Research Project

Project/Area Number 60870062
Research Category

Grant-in-Aid for Developmental Scientific Research

Allocation TypeSingle-year Grants
Research Field Ophthalmology
Research InstitutionJichi Medical School

Principal Investigator

OHARA Kunitoshi  Jichi Medical School, 医学部, 助教授 (00137704)

Co-Investigator(Kenkyū-buntansha) 阿部 国臣  甲南カメラ研究所, 製品部, 部長代理
ABE Kuniomi  Head of production
Project Period (FY) 1985 – 1986
Project Status Completed (Fiscal Year 1986)
Budget Amount *help
¥11,000,000 (Direct Cost: ¥11,000,000)
Fiscal Year 1986: ¥3,000,000 (Direct Cost: ¥3,000,000)
Fiscal Year 1985: ¥8,000,000 (Direct Cost: ¥8,000,000)
KeywordsIntraocular lenses / Specular microscopy / Optics / Furrow-like defects
Research Abstract

We aimed to make the full-automatic measurement system for computation of the corneal endothelial cell area, but we had difficulties that it was impossible to measure the corneal endothelial cells in vivo in various conditions in real time automatically because the each cell image in the specular microscope was not sharp and had a low contrast, and the proceessing speed of a microcomputer was still slow. Therefore, we developed a semi-automatic system for measurement of the corneal endothelial cell area, cell density, coefficient of variation, and distribution of cell area using a specular microscope (Konan), a TV camera, an image processing unit including a digitizer, and a microcomputer. In this system, the corneal endothelial cell image was obtained from the TV camera and the image processor connected to the specular microscope, and were digitized manually using 2 midpoints of confronting cell sides of 40 cell sample. We applied this system to measurement of various sizes of the endothelial cells and data was compared with those which had been computed by the conventional digitizing system. This system required 2 minutes to compute the data. The error of mean endothelial cell area in this system was within 10%. This system in this specifications was very useful to know the status of the corneal endothelial cell morphology, and could be used in specular microscopy of the endothelial cells in clinical practice.

Report

(2 results)
  • 1986 Final Research Report Summary
  • 1985 Annual Research Report
  • Research Products

    (2 results)

All Other

All Publications (2 results)

  • [Publications] 笠原達也,阿部国臣,伊野田繁,大原國俊: 日本眼科紀要. 38. (1987)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1986 Final Research Report Summary
  • [Publications] Tatsuya Kasahara, Kuniomi Abe, Shigeru Inoda, M.D. and Kunitoshi Ohara, M.D.: "A new system to quantify the corneal endothelial morphology for clinical use." Folia Ophthalmologica Japonica. 38,No.6,. (1987)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1986 Final Research Report Summary

URL: 

Published: 1987-03-31   Modified: 2016-04-21  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi