• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to previous page

Development of the Portable negative pressure ventilator

Research Project

Project/Area Number 61870055
Research Category

Grant-in-Aid for Developmental Scientific Research

Allocation TypeSingle-year Grants
Research Field Thoracic surgery
Research InstitutionKyoto University

Principal Investigator

HITOMI Shigeki M.D.Professor  Department of Thoracic Surgery, Chest Disease Research Institute, 胸部疾患研究所, 教授 (80173186)

Co-Investigator(Kenkyū-buntansha) 野口 康夫  住ベメディカル, 主任研究員
CHIHARA Koji M.D.Assista  Department of Thoracic Surgery, Chest Disease Research Institute, 胸部疾患研究所, 助手 (70197614)
TAMURA Koich M.D.Associa  Department of Artificial Organs, Chest Disease Research Institute, 胸部疾患研究所, 助教授 (50144390)
WADA Hiromi M.D.Associat  Department of Thoracic Surgery, Chest Disease Research Institute, 胸部疾患研究所, 助教授 (90167205)
SHIMIZU Yasuhiko M.D.Pro  Department of Experimental Surgery, Research Center for Medical Polymers, 医用高分子センター, 教授 (00027111)
NOGUCHI Yasuo  S.B. Medical Company
渡部 智  京都大学, 医用高分子研究センター, 助教授 (40167127)
Project Period (FY) 1986 – 1988
Project Status Completed (Fiscal Year 1988)
Budget Amount *help
¥12,700,000 (Direct Cost: ¥12,700,000)
Fiscal Year 1988: ¥500,000 (Direct Cost: ¥500,000)
Fiscal Year 1987: ¥4,000,000 (Direct Cost: ¥4,000,000)
Fiscal Year 1986: ¥8,200,000 (Direct Cost: ¥8,200,000)
Keywordscuirass respirator / negative pressure ventilator / assisted ventilat muscle fatigue / respiration sensor / 呼吸筋疲労 / Cuirass レスピレータ / 呼吸センサー / Cuirassレスピレータ / 陰圧式人工呼吸器 / Cuirass respirator
Research Abstract

A new cuirass respirator which synchronizes with spontaneous respitation and operates in various ventilation assistance modes has been developed and applied to patients wth acute and chronic respiratory failure.
Synchronization of the respirator was made possible by analysis of temperature changes in the nasal cavity or in the artificial airway during spontaneous respiration with the use of a nasal thermistor with a differential circuit or a newly developed pyroelectric heat sensor.
The respirator can provide not only respiration synchronized mode but also controlled respiration mode. In the latter, I/E ratio is variable. In both modes, negative-zero mode (inspiratory support) and negative-zero-positive-zero mode (inspiratory-expiratory support) ar able to be set. The ratio of assistance frequency to spontaneous respiration can be set to 1:1,1:2 and 1:3. Furthemore, a back-up mode maintain adequate ventilation when the patient fall to hypopnea or apnea furing synchronized mode.
Thirty one … More patients with thoracic restrictive disease, neuromuscular disease,COPD and others were assisted by the cuirass respirator for periods 1 to 6 hours during the daytime intermittently.
Respiratory assitance relieved dyspnea in all patients with acute respiratory failure. Herat rate and respiratory rate decreased after the assistance. Arterial blood gas analyses before and after respiratory assistance showed improvement. It showed the advantage of the respirator that in some patients good effects lasted after using.
In experimental study with the use of mongeral dogs, pulmonary artery pressure decreased during assisted negative pressure ventilation by the cuirass respirator suspected a decrease of after load for right ventricule.
Impairments of circulatory status were not induced during assited negative pressure ventilation by the cuirass respirator in patients who underwent thoracic surgery.
We conclude that the new cuirass respirator, which can synchronize with spontaneous respiration, provides effective assistance for patients with respiratory failure and stabilizes thier respiratory status. Less

Report

(3 results)
  • 1988 Annual Research Report   Final Research Report Summary
  • 1986 Annual Research Report
  • Research Products

    (17 results)

All Other

All Publications (17 results)

  • [Publications] 千原幸司: 人工臓器. 16. 611-615 (1987)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] 千原幸司: 病態生理. 6. 958-964 (1987)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] 河原崎茂孝: 人工臓器. 17. 783-786 (1988)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] 千原幸司: 人工呼吸. 5. 128-134 (1988)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] K.Chigara: Chest. 94. 44S (1988)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] 人見滋樹: 胸部外科. (1989)

    • Description
      「研究成果報告書概要(和文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] Chihara, K; Kawarasaki, S; Shimizu, Y; Isobe, J; Terada, Y; Matsunobe,S; Mizuno, H; Hitomi, S; Yagata, K; Noguchi, Y; Mikami, T: "Development of negative pressure ventilator" The Japanese Journal of Artifical Organs. 16. 611-615 (1987)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] Chihara, K; Kawarasaki, S; Hitomi, S; Shimizu, Y: "Development of a portable respirator." Byoutaiseiri. 6. 958-964 (1987)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] Chihara, K,; Kawarasaki, S.; Hitomo, S.; Shimizu-,: "A new cuirass respirator synchronizing with spontaneous respiration." Mechanical Ventilation. 5. 128-134 (1988)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] Chihara, K.; Kawarasaki, S.; Hitomi, S.; Shimizu, Y.: "A new cuirass respirator synchronizing with spontaneous respiration and its clinacal application." Chest. 94. 44 (1988)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] Kawarasaki, S.; Chihara, K.; Hitomi, S.; Shimizu, Y.; Noguchi, Y.; Yagata, K.: "A new respiratory sensor using with pyroelectric sensor" The Japanese Journal of Artificial Organs. 17. 783-786 (1988)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] Hitomi, S.; Chihara, K.; Kawarasaki, S.; Kobayashi, J.; Shimizu, Y.; Kobayashi, J.: "Right heart assistance for patients following lunt resection." Thoracic Surgery. (1989)

    • Description
      「研究成果報告書概要(欧文)」より
    • Related Report
      1988 Final Research Report Summary
  • [Publications] 河原崎茂孝: 人工臓器. 17. 783-786 (1988)

    • Related Report
      1988 Annual Research Report
  • [Publications] 千原幸司: 人工呼吸. 5. 128-134 (1988)

    • Related Report
      1988 Annual Research Report
  • [Publications] K.Chihara: Chest. 94. 44S (1988)

    • Related Report
      1988 Annual Research Report
  • [Publications] 人見滋樹: 胸部外科.

    • Related Report
      1988 Annual Research Report
  • [Publications] 千原幸司: 人工臓器. 16. 611-615 (1987)

    • Related Report
      1986 Annual Research Report

URL: 

Published: 1987-03-31   Modified: 2016-04-21  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi