1993 Fiscal Year Final Research Report Summary

Research of Totally Implantable Linear Pulse Motor-Driven Artificial Heart

Research Project

Project/Area Number 03405005
Research Category

Grant-in-Aid for General Scientific Research (A)

Allocation TypeSingle-year Grants
Research Field 広領域
Research InstitutionShinshu University

Principal Investigator

YAMADA Hajime  Shinsyu Univ., Engineering, Professor, 工学部, 教授 (70020972)

Co-Investigator(Kenkyū-buntansha) YAMAMOTO Y.  Nagano National College, Professor, 教授 (30043029)
MORIMOTO M.  Shinsyu Univ., Medicine, Associate Professor, 医学部, 助教授 (10020885)
WAKIWAKI H.  Shinsyu Univ., Engineering Associate Professor, 工学部, 助教授 (50240461)
Project Period (FY) 1991 – 1993
KeywordsArtificial heart / Mock testing / Electromagnetic actuator / Linear motor / Linear pulse motor / Animal testing / Permendur / Blood pump
Research Abstract

Totally lmplantable Linear-type Artificial Heart under Animal Testing
Our group has been considering the possibility of using a linear pulse motor (LPM) as a driving source, and has developed a series of artificial heart models with the motor
The newly developed implantable linear-type artificial heart (LAH) has two blood pumps, which are inflated and deflated by the LPM alternately, causing a pusher plates to pulsate the blood
The LAH measures 85 mm in diameter * 95mm wide (0.55 litters total volume) , and weighs 1.2kg. The LPM comprises a flat mover as the secondary member and two stators as the primary member. Two linear bearings are used to support them. Both stators and mover are made of laminated thin layrs of Permendur with high saturation flux density.
The LPM provides a kinetic thrust of 50-60 newtons at 60-100 beats per min., the speed required by an artificial heart
A Windkessel-type mock circulatory system was used to evaluate the performance of the LAH.The results were good enough to meet the flow rate requirements of a human at rest. Maximum aortic pressure (AoP max) of 141 mmHg, maximum pulmonary arterial pressure (PAP max) of 28 mmHg, and maximum flow rate of 5.1 liters per min. were obtained
In collaboration with Hiroshima University Medical School, animal testing using a sheep was conducted on September 24 1993. The experiment confimed that the implanted artificial heart could function as expected in a living body. The heart worked properly for up to two hours
The development profect still faces a number of unsolved problems before device can be applied clinically, but it has been steadily advancing toward the final target

  • Research Products

    (16 results)

All Other

All Publications (16 results)

  • [Publications] 山田一,他5: "リニアパルスモータを使用した大動脈バルーンポンプ" 電気学会論文誌D. 113-D. 532-538 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 山口昌樹,他3: "リニアモータ駆動型完全人工心臓2機種の性能比較" 日本応用磁気学会誌. 17. 453-458 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] M.Yamaguchi,et al: "Performance test of a linear pulse motor-driven artificial heart" IEEE Trans.J.Magnetics Japan. 8. 130-135 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] M.Yamaguchi,et al: "Performance evaluation of a linear electromagnetic solenoid for a total artificial heart" 1993 Annual Conference of IEE Japan-Industry Applications Society-. J93. 97-100 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] H.Yamada: "Totally implantable linear-type artificial heart under animal testing" MEJ(Medical Equipment Journal of Japan). 37. 5,21 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 山口昌樹,山田一: "完全人工心臓駆動用リニアパルスモータの信頼性評価" 電気学会論文誌D. 114-D. 94-100 (1994)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] T.Akutsu,H.Koyanagi Edt: "Heart Replacement-Artificial Heart4-" Springer-Verlag, 424 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 尾前照雄 編: "厚生省循環器病 研究報告集" 国立循環器病センター, 647 (1993)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] H.Yamada, H.Wakiwaka, T.Yano, A.Watarai, M.Karita, and M.Morimoto: "Intra-aortic balloon pump using linear pulse motor" T.IEE Japan. Vol.113-D No.4. 532-538 (1993)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] M.Yamaguchi, K.Kobayashi, Y.Yamamoto, and H.Yamada: "Performance comparison of two linear motor-driven total artificial hearts" Journal on Magnetics in Japan. Vol.17, No.2. 453-458 (1993)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] M.Yamaguchi, T.Yano, M.Karita, Y.Yamamoto, S.Yamada, and H.Yamada: "Performance test of a linear pulse motor-driven artificial heart" IEEE Trans. J.on Magnetics in Japan. Vol.8, No.2. 130-136 (1993)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] M.Yamaguchi, K.Kobayashi, and H.Yamada: "Performance evaluation of a linear electromagnetic solenoid for a total artificial heart" 1993 Annual Conference of IEE Japan - Industry Applications Society. J.93-20. 97-100 (1993)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] H.Yamada: "Totally implantable linear-type artificial heart under animal testing" MEJ (Medical Equipment Joumal of Japan). Vol.37, No.12. 5,21 (1993)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] M.Yamaguchi, and H.Yamada: "Reliability evaluation of a linear pulse motor for total artificial heart driving" T.IEE Japan. Vol.114-D, No.1. 94-100 (1994)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] T.Akutsu, and H.Koyanagi Edited: Heart Replacement -Artificial Heart 4 -. Springer - Verlag Tokyo, 424 (1993)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] T.Ogata Edited: Research Results for Cardiovascular Disease of the Ministry of Healty and Welfare. National Cardiovascular Center, 647 (1993)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 1995-03-27  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi