• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

1995 Fiscal Year Final Research Report Summary

Development of a Simple Monitoring Device for Ultra-Trace Silicate lon in Quality Control High-Purity Water Used in Semiconductor Production.

Research Project

Project/Area Number 06555252
Research Category

Grant-in-Aid for Developmental Scientific Research (B)

Allocation TypeSingle-year Grants
Research Field 工業分析化学
Research InstitutionTohoku University

Principal Investigator

HOSHINO Hitoshi  Tohoku University, Faculty of Engineering, Associate Professor., 工学部, 助教授 (20124620)

Co-Investigator(Kenkyū-buntansha) SUGAWARA Katsuyoshi  Miyagi Oki Elect. Co. Manager, 係長
KOBAYASHI Setsu  Miyagi Prefectural Institute of Technology, Chief Scientist, 総括研究員
KONNO Shinichi  Ihara Rinken Ltd. Manager, 係長
Project Period (FY) 1994 – 1995
KeywordsSilicate / Silica / High-Purity Water / Quality Control / Monitoring
Research Abstract

The aim of this project is to develop a simple monitoring device for quantitation of ultra-trace (sub-ppb to ppb) silicate ion in high-purity water using flow preconcentration-spectrophotometric detection system. The basic ideas and the underlying chemistry are :
1.Silicate ion is converted to molybdosilicate blue which has the absorption maximum in near infrared region, at 830nm. Thus, one can devise a cheap detector based on the optosemiconductors, such as LEDs and photodiodes (PD).
2.Anionic molybdosilicate blue is adsorbed on cellulose-type materials as the ion-pair with a large organic cation which is very readily desorbed (eluted) with certain organic solvents. These leads the high-performance flow preconcentration systems.
The optimized system consists of two pump units for sampling and elution, respectively, and a short column (4 cm) packed with ashless pulp. Molybdosilicate blue in 30 ml is concentrated on the ashless pulp column as an ion-pair with octyltrimethyl ammonium ion and the ion-pair is eluted with 80 wt% acetonitrile-water mixture down to the LED-PD detector operated at 830nm.
In order to ensure the practical capability, the prototype equipment is now under field testing in some semiconductor production manufacturers and electric power plants.

  • Research Products

    (2 results)

All Other

All Publications (2 results)

  • [Publications] 星野 仁: "極微量ケイ酸イオン計測のための前段濃縮-吸光検出装置の開発研究" みやぎ工業技術情報. 137号. 2-4 (1996)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] HOSHINO,Hitoshi: "Study of Preconcentration-Photometric Detection System for the Monitoring of Ultra-Trace Silicate Ion in High-Purity Water." Miyagi Technical Information. vol.137. 2-4 (1996)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 1997-03-04  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI