• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

1997 Fiscal Year Final Research Report Summary

Development of Biomimetic Prosthetic Hand

Research Project

Project/Area Number 07555431
Research Category

Grant-in-Aid for Scientific Research (A)

Allocation TypeSingle-year Grants
Section展開研究
Research Field 計測・制御工学
Research InstitutionKobe University

Principal Investigator

AKAZAWA Kenzo  Kobe University, Faculty of Engineering, Professor, 工学部, 教授 (30029277)

Co-Investigator(Kenkyū-buntansha) OKUNO Ryuhei  Kobe University, Faculty of Engineering, Research Associate, 工学部, 助手 (90294199)
UCHIYAMA Takanori  Keio University, Faculty of Science and Technology, Research Associate, 理工学部, 助手 (50243324)
YOSHIDA Masaki  Kobe University, Faculty of Engineering, Associate Professor, 工学部, 助教授 (30174949)
Project Period (FY) 1995 – 1997
Keywordsmyoelectric prosthetic hand / neuromuscular control system / biomimetic / sensory feedback / vibrotactile stimuli / simulator
Research Abstract

The purpose of this study was to develop a biomimetic prosthetic hand controlled by electromyogram. This hand mimicked dynamic properties of human neuromuscular control system and had a supplementary sensory feedback system.
We developed a portable control system of the biomimetic hand. This system consisted of processing units of EMG signals, a digital servo system of DC motor and two sets of battery. A new type of flexible glove for the biomimetic hand was developed. This rubber glove was made of highly deproteinized natural rubber with low deformation resistance and did not disturb a motion of the prosthetic hand to a large extent.
We developed two types of supplementary sensory feedback systems, which transmitted information of the force applied to the finger of the prosthesis to its user with vibotactile stimuli. In one system, optimum frequency (250Hz) of vibration was utilized. At this frequency, the amplitude of stimuli that the user could sense was smallest. Other system utilized infrasonic vibrotactile stimuli. Utility of these systems were shown by the experiment where a normal subject was grasping soft objects with the prosthetic hand.
We developed a simulator of myoelectric hand with which amputees could try to operate the myoelectric hand and to decide appropriate parameters of the control system. In this simulator, the motion of the prosthetic hand and posture of user's upper extremity were displayd with real-time 3D computer graphics. Myoelectric control experiments showed that the user was able to control finger-angles of prosthetic hand and to grasp the object virtually with this simulator.

  • Research Products

    (12 results)

All Other

All Publications (12 results)

  • [Publications] Ryuhei OKUNO et al: "Development of Biomimetic Prosthetic Hand Controlled by Electromyogram" Proc.4th Int.Workshop AMC. 1. 103-108 (1996)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 奥野 竜平 他: "皮膚振動刺激による筋電義手感覚フィードバック装置の検討" 電気学会論文誌. 116-C. 1246-1251 (1996)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 奥野 竜平 他: "可聴周波数以下の皮膚振動刺激を用いた義手感覚フィードバック装置" 電子情報通信学会技術報告書. MBE96-137. 99-106 (1997)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Kenzo Akazaw et al: "Biomimetic EMG-Prosthesis Hand" Proc.18th Ann.Int.Conf.IEEE EMBS. 258-259 (1996)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 木村 貴俊: "MYOBOCK筋電義手の3DCGシミュレータの開発" 電子情報通信学会技術報告書. MBE97-90. 29-34 (1997)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 川崎 賢一郎 他: "柔軟性の高い筋電義手グローブの試作" 電子情報通信学会技術報告書. MBE97-141. 109-115 (1998)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Ryuhei OKUNO et al: "Development of Biomimetic Prosthetic Hand Controlled by Electromyogram" Proc. 4th Int.Workshop AMC. 1. 103-108 (1996)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Kenzo AKAZAWA et al: "Biomimetic EMG-Prosthesis Hand" Proc. Annb.1 Int.Conf.IEEE EMBS. 258-259 (1996)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Ryuhei OKUNO et al: "Basic Study of vibrotactile sensory feedback system for biomimetic powered-prosthetic hand" Trans.of IEE of Japan. 116-C. 1246-1251 (1996)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Ryuhei OKUNO et al: "Infrasonic vibrotactile sensory feedback system for the myoelectric hand prosthesis" Technical Report of IEICE. MBE96-137. 99-106 (1997)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Takatoshi KIMURA et al: "3D-CGSimulator of Myoelectric Hand" Technical Report of IEICE. MBE97-90. 29-34 (1997)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Kenichiro KAWASAKI et al: "Study of low deformation resistance rubber-glove for the biomimetic hand" Technical Report of IEICE. MBE97-141. 109-115 (1998)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 1999-03-16  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI