Basic research concerning control method of bi-ventricular assist with nonpulsatile flow pump

Research Project

Project/Area Number 11557102
Research Category

Grant-in-Aid for Scientific Research (B)

Allocation TypeSingle-year Grants
Section展開研究
Research Field Thoracic surgery
Research InstitutionNational Cardiovascular Center Research Institute

Principal Investigator

TAKANO Hisateru  National Cardiovascular Center Research Institute, Deputy, 副所長 (60028595)

Co-Investigator(Kenkyū-buntansha) NISHINAKA Tomohiro  National Cardiovascular Center Research Institute, Department of Artificial Organs, Laboratory Chief, 人工臓器部, 室長 (00256570)
TATSUMI Eisuke  National Cardiovascular Center Research Institute, Department of Artificial Organs, Laboratory Chief, 人工臓器部, 室長 (00216996)
TAENAKA Yoshiyuki  National Cardiovascular Center Research Institute, Department of Artificial Organs, Director, 人工臓器部, 部長 (00142183)
TSUKIYA Tomonori  National Cardiovascular Center Research Institute, Department of Artificial Organs, Staff, 人工臓器部, 室員 (00311449)
Project Period (FY) 1999 – 2001
KeywordsArtificial heart / nonpulsatile flow pump / Continuous flow / biventricular support / Control / Motor Current
Research Abstract

Mechanical circulatory support is becoming an important therapeutic option both as a bridge to heart transplantation and for permanent use. Several types of nonpulsatile flow pumps are being developed for long-term circulatory support, having the advantage of compactness and low cost as compared to pulsatile pumps. The pathophysioloey of long-term nonpulsatile ventricular support, however, is not yet clearly understoocL The control method of continuous flow pumps especially in bi-ventncular assist with nonpulsatile flow pump has not been established. We have attempted basic research concerning control method of bi-ventricular assist with nonpulsatile flow pump. The hemodynamics under ventricular assist with continuous flow pump is quite particular including reduction in arterial pulsatility. We firstly evaluated physiological changes due to ventricular assist with the nonpusatile flow pump. We found that the diminution of systemic yasoconstrictive function and progression of cardiac parasympathetic nervous activity. However no obvious changes were not observed in central nervous system, baroreceptor function, hormonal factors, and pulmonary system. The control tactics for the optimized operation of a continuous-flow VAD using the driving motor current has been proposed. The good correlation between the motor current and flow rate allowed us to detect adverse pumping condition such as inflow cannulae sucking and even to estimate assist status by the VAD. The control of the continuous-flow pumps for biventricular support has also been examined. Maintenance of the circulatory balance between the left and the right pump was attempted by adjusting the hydrodynamic characteristics of the rignt pump. Use of additional resistance of the outflow tubing could successfully reduce the flow rate fluctuation caused by the change in circulatory resistance. The flow rate of the right pump was stabilized by this device.

  • Research Products

    (12 results)

All Other

All Publications (12 results)

  • [Publications] Tsukiya T., Taenaka Y., Nishinaka T.et al.: "Application of indirect flow rate measurement using motor driving with an integrated motor"Artificial Organs. 25. 692-696 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Tsukiya T., Taenaka Y., Tatsumi E.et al.: "Performance of newly developed implantable centrifugal blood pump"ASAIO Journal. 47. 559-562 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Nishinaka T., Tatsumi E., Nishimura T.et al.: "Change in vasoconstrictive function during prolonged non pulsatile left heart bypass"Artificial Organs. 25. 371-375 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Nishinaka T., Tatsumi E., Nishimura T.et al.: "Effects of reduced pulse pressure to the cerebral metabolism during prolonged nonpulsatile left heart bypass"Artificial Organs. 24. 676-679 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Nishinaka T., Tatsumi E., Nishimura T.et al.: "Cardiac autonomic nervous function during long-term nonpulsatile left heart bypass"Artificial Organs. 23. 500-503 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Nishimura T., Tatsumi E., Nishinaka T.et al.: "Prolonged nonpulsatile left heart bypass diminishes vascular comtractility"Int J Artif Organs. 22. 492-498 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Nishimura T., Tatsumi E., Nishinaka T. et al: "Prolonged nonpulsatile left heart bypass diminishes vascular comtractility"Int J Artif Organs. 22. 492-498 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Nishinaka T., Tatsumi E., Nishimura T. et al: "Cardiac autonomic nervous function during long-term nonpulsatile left heart bypass"Artificial Organs. 23. 500-503 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Nishinaka T., Tatsumi E., Nishimura T. et al: "Effects of reduced pulse pressure to the cerebral metabolism during prolonged nonpulsatile left heart bypass"Artifi Organs. 24. 676-679 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Nishinaka T., Tatsumi E., Nishimura T. et al: "Change in vasoconstrictive function during prolonged non pulsatile left heart bypass"Artificial Organs. 25. 371-375 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Tsukiya T., Taenaka Y., Tatsumi E. et al: "Performance of newly developed implantable centrifugal blood pump"ASAIO Journal. 47. 559-562 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Tsukiya T., Taenaka Y., Nishinaka T. et al: "Application of indirect flow rate measurement using motor driving signals to a centrifugal blood pump with an integrated motor"Artif Organ. 25. 692-696 (2001)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 2003-09-17  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi