Micro gas analytical system by means of radial electrophoresis pump

Research Project

Project/Area Number 11640611
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field 分離・精製・検出法
Research InstitutionKumamoto University

Principal Investigator

TODA Kei  Kumamoto University, Faculty of Science, Associate professor, 理学部, 助教授 (90264275)

Project Period (FY) 1999 – 2001
KeywordsELECTROPHORESIS PUMP / GAS ANALYSIS / DROPLET MICRORING ELECTRODE / HYDROGEN SULFIDE / 二酸化硫黄
Research Abstract

In this research, a simple liquid flow method, which could be used in the field, was developed and it was applied to monitoring of atmospheric pollutants. First, a new concept of flowing method "radial electrophoresis pump" was designed and investigated. In this method, gas absorbing solution is flowed electrophoretically with only several volts of voltage. In the beginning of the test it seemed that it worked, but actually the liquid was moved by bubbles generated by the electrodecomposition of water. Then, other methods were tested ; the absorbing solution could be flowed through 10-cm micro channel in a glass chip and same length silicone tubing with several tens of volts. Next, gravity method and compressing method were also tested and they worked well especially compressing with a small compressor and pressure sensor. These flow methods were applied to some kinds of gas analysis. One is a droplet ring electrode fabricated on a chip of micro capillary This device showed very fast response compared to the conventional chronoamperometry The compressing method was applied to a high sensitive H_2S determination system that was composed of tubular gas diffusion scrubber and solid state photo devices. Very low level of H_2S even back ground level could be measured continuously in this method, because stable liquid flow without pulsation was obtained. This system could be operated by battery and could be used for several days in outside field without AC power source. The results obtained with the system showed a good correlation with a conventional tape meter, and further, the time and concentration resolutions were much better in this system.

  • Research Products

    (10 results)

All Other

All Publications (10 results)

  • [Publications] 戸田 敬: "亜硫酸ガスの高感度マイクロ検出器"硫酸と工業. 52. 1-6 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Toda: "A wall-jet ring disk electrode fabricated within a thin-layered micromachined cell"Journal of Electroanalytical Chemistry. 479. 57-63 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 戸田 敬: "液的マイクロリング電極による電気化学的ガス計測方法の検討"分析化学. 49. 989-995 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Toda: "Fluorometric field instrument for continuous measurement of atmospheric hydrogen sulfide"Analytical Chemistry. 73. 5716-5724 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Toda: "Measurement of atmospheric hydrogen sulfide by continuous flow fluorometry"Analytical Sciences. 18(in press). (2002)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K. Toda, H. Inoue, S. Ohira and I. Sanemasa: "High sensitive micro detector for measurement of sulfur dicxide"Ryusan to Kogyo. 52. 1-6 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K. Toda, S. Oguni, Y. Takamatsu and I. Sanemasa: "A wall-jet ring disk electrode fabricated within a thin-layered micromachined cell"Journal of Electroanalytical Chemistry. 479. 57-63 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K. Toda, M. Kitazaki and I. Sanemasa: "Amperometric gas detection using a micro-ring electrode coupled with a gas-absorbing droplet"Bunseki Kagaku. 49. 989-995 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K. Toda, P.K. Dasgupta, J. Li, G.A. Tarver and G.M. Zarus: "Fluorometric field instrument for continuous measurement of atmospheric hydrogen sulfide"Analytical Chemistry. 73. 5716-5724 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K. Toda, P.K. Dasgupta, J. Li, G.A. Taver, G.M. Zarus, S. Ohira and E.L. Loree: "Measurement of atmospheric hydrogen aulfide hy continuous flow Fluorometry"Analytical Sciences. (in press).

    • Description
      「研究成果報告書概要(欧文)」より