• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2000 Fiscal Year Final Research Report Summary

STUDY OF WAVEFORM ANALYSIS WITH A 3-D VECTOR MEASUREMENT OF THE MEG.

Research Project

Project/Area Number 11680850
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field Biomedical engineering/Biological material science
Research InstitutionTOKYO DENKI UNIVERSITY

Principal Investigator

UCHIKAWA Yoshinori  COLLEGE OF SCIENCE AND ENGINEERING, PROFESSOR, 理工学部, 教授 (90147455)

Co-Investigator(Kenkyū-buntansha) KOBAYASHI Koichiro  APPLIED SUPERCONDUCTIVITY RESEARCH LABORATORY, INSTRUCTOR, 超電導応用研究所, 助手 (60277233)
Project Period (FY) 1999 – 2000
Keywordsmagnenetoencephalogram / MEG / 3-D vector maesurement / inverse problem / multiple source estimation / singular value decomposition / time-frequency analysis
Research Abstract

In the research on the magnetoencephalogaram (MEG) measurement the magnetic field perpendicular to the scalp is widely used. There are problems of separating multiple sources overlapping in time when many distinct areas of the cortex are active. In these cases the magnetic field distribution of normal component is not helpful for estimating the location and number of sources. owing to the lack of a dipole pattern. From viewponit of waveform analysis, it is also desired a certain method to detect the transient weak signal hid by the intense signal. Concerning to overcom the previous ploblems, new methodology is necessary to develop in both of not only signal processing but also SQUID instrumentation. First, we have developed a three dimensional (3-D) second-order gradiometer connected to three superconducting quantum interference devices (SQUIDs) for vector measurement of the MEG that can detect magnetic field components perpendicular and tangential to the scalp simultaneously. Second, we proposed a new method of source estimation algorithm using spatio-temporal data of 3-D MEG and singular value decomposition analysis. In this algorithm the number of sources are decided by number of dominant singular value of spatio-temporal 3-D MEG data. The source localization is done as pattern matching by signal subspace of singular vectors and projection of an estimating source.This proposed method can estimate two and/or three sources overlapping in time, which are 2 cm apart and 15 degree in phase difference. effectively.

  • Research Products

    (24 results)

All Other

All Publications (24 results)

  • [Publications] 小林宏一郎: "三次元ベクトル計測に基づく心磁図の時間・周波数解析"日本応用磁気学会誌. 25(末定). (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 小林宏一郎: "三次元心磁図計測に基づく伝導障害患者の波形解析"日本応用磁気学会誌. 25(末定). (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] M.Liu: "Analysis of high frequency components of the QRS complex in α 3-Q MCG using wavelet packets."日本応用磁気学会誌. 25(末定). (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 小林宏一郎: "三次元脳磁界計測に基づく特異値分解を用いた複数信号源推定の検討"電気学会論文誌A. 120-A. 823-828 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 小林宏一郎: "P-R分節心磁図の三次元ベクトル計測"日本応用磁気学会誌. 24. 939-942 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] M.Liu: "Discussion of detecting high frequency components of the QRS complex in 3-D MCG using wavelet transform"日本応用磁気学会誌. 24. 943-946 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Sai: "MCG measurement using a high-Tc SQVID in patient with inferior infarction"日本応用磁気学会誌. 24. 947-950 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Sai: "Analysis of MCG measured by high-Tc SQUID with spatial filter method"Trans.IEE of Japan. 120-C. 1201-1207 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 川勝真喜: "全頭型ベクトルSQUID磁界計測システムのキャリブレーションとポジショニング"日本生体磁気学会論文誌. 12. 3-9 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Koichro Kobayashi: "Frequency analysis of premature ventricular contraction using 3-D MCG measurement"IEEE Trans.Magnetics. 35. 4112-4114 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] H.Ohta: "Neuromagnetic SQUID measurement in a superconducting magnetic measurement"IEEE Trans.Applied Superconductivity. 9. 4073-4076 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] H.Ohta: "A 64-channel whole-head SQUID system in a superconducting magnetic shield"Supercond.Soci.Technol.. 12. 762-765 (1999)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] K.Kobayashi: "Time-Frequency Analysis of MCGs based on 3D vector measurement"J.Magn.Soc.Japan. Vol.25(in press). No.4-2 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Kobayashi: "Waveform analysis of conduction disease based on 3-D MCG measurements"J.Magn.Soc.Japan. Vol.25(in press). No.4-2 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] M.Liu: "Analysis of high frequency components of QRS complex in a 3-D MCG using wavelet packets"J.Magn.Soc.Japan. Vol.25(in press). No.4-2 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Kobayashi: "Multiple source estimation using singular value decomposition on a three dimensional measurement of magnetoencephalogram"Trans.IEE of Japan. Vol.120-A No.8/9. 823-828 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Sai: "Analysis of MCG measured by high-Tc SQUID with spatial filter method"Trans.IEE of Japan. Vol.120-C, No.8/9. 1201-1207 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Kobayashi: "3-D MCG vector measurement of the PR segment"J.Magn.Soc.Japan. Vol.24, No.4-2. 939-942 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] M.Liu: "Discussion of detecting high frequency components of the QRS complex in 3-D MCG using wavelet transform"J.Magn.Soc.Japan. Vol.24, No.4-2. 943-946 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] K.Sai: "MCG measurement using a high-Tc SQUID in patient with inferior infarction"J.Magn.Soc.Japan. Vol.24, No.4-2. 947-950 (2000)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Masaki Kawakatsu: "A calibration of the sensitivity and position of the sensors for whole-head vector SQUID magnetoencephalography sysytem and positioning sysytem for the sysytem"The Journal of Japan Biomagnetism and Bioelectromagnetics Society. Vol.12, No.2. 3-9 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] H.Ohta: "Neuromagnetic SQUID measurement in a superconducting magnetic shield"IEEE Trans.Applied Superconductivity. Vol.9, No.2. 4073-4076 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] H.Ohta: "A 64-channel whole-head SQUID system in a superconducting magnetic shield"Supercond.Soci.Technol. Vol.12. 762-765 (1999)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Koichiro Kobayashi: "Frequency analysis of premature ventricular contraction using 3-D MCG measurement"IEEE Trans.Magnetics. Vol.35, No.5. 4112-4114 (1999)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 2002-03-26  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI