2001 Fiscal Year Final Research Report Summary

Development of Ultrasonically Activated Bending Scalpel for Endoscopic Surgery

Research Project

Project/Area Number 12555072
Research Category

Grant-in-Aid for Scientific Research (B)

Allocation TypeSingle-year Grants
Section展開研究
Research Field Intelligent mechanics/Mechanical systems
Research InstitutionShinshu University

Principal Investigator

HASHIMOTO Minoru  Shinshu University, Faculty of Textile Science and Technology, Professor, 繊維学部, 教授 (60156297)

Co-Investigator(Kenkyū-buntansha) YAMAUCHI Kiyoshi  Tokin Co., Director, 材料開発研究所, 研究部長
NAKAMURA Yoshihiko  University of Tokyo, Faculty of Engineering, Professor, 大学院・工学系研究科, 教授 (20159073)
YOSHINAKA Heiji  Kagoshima University, School of Medicine, Associate Professor, 医学部・附属病院, 助教授 (80191625)
Project Period (FY) 2000 – 2001
KeywordsEndoscopic Surgery / Ultrasonically activated Scalpel / Super Elasticity Alloy / Minimally Invasive Surgery / Medical Robot / Laparoscopic Sergery
Research Abstract

Ultrasonically activated scalpels are used clinically to provide hemostatic cutting in endoscopic surgery. To assist the physician during endoscopic surgery, we developed an ultrasonically activated bending scalpel constructed of super elasticity alloy ( SEA in short ). First, we performed the vibration analysis and its verification in bending state of ultrasonically activated scalpel. We found out that longitudinal vibration was transmitted even if probe was bent. The influence of transverse vibration is very small. There is no trouble destroying tissue contacts with the probe. Next, a SEA bending probe and its bending angle control system were manufactured as an experiment. From the experiment using chicken, we evaluated the validity of the bending scalpel. From this research, we found out that it is possible to bend the probe of ultrasonically activated scalpel by using Super Elasticity Alloy, and the activation of ultrasonically activated scalpel could be realized.
Although, a lot of surgeons have used ultrasonic coagulating and cutting systems and shears ( LCS ) in performing endoscopic thryroidectomies and parathyroidectomics and parathyroidectomies, our experience showed that some patients had temporary paralysis of the recurrent laryngeal nerve ( RLN ) after these operations. We sometimes noticed that the heat of the blade of the LCS was raised just after use; therefore, we designed the RLN model and investigated the damage to the RLN attributable to by heat of LCS. We investigated the effects of the heat of the blade of LCS upon rats' femoral and sciatic nerves using temperature measurements, histological examinations and evoked electromyography. Our results suggest that we should not touch the RLN directly with the blade just after using it, and that it is possible to use LCS at a distance of 3mm from the RLN for less than 20 sec at level 3. In order to maintain these distances requires visualizing the RLN in the view of endoscope at the neck surgery.

  • Research Products

    (11 results)

All Other

All Publications (11 results)

  • [Publications] 若子倫菜: "内視鏡下外科手術用超音波駆動メスの湾曲化 第2報 基本構想と振動状態の評価"第19回日本ロボット学会学術講演会. 955-956 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 若子倫菜: "内視鏡下手術のための能動湾曲超音波メスの開発"日本機械学会ロボティクス・メカトロニクス講演会予稿集. (発表予定). (2002)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Rina Wakako: "Development of Ultrasonically Activated Bending Scalpel for Endoscopic Surgery"2002 IEEE/RSJ International Conference on Intelligent Robots and Systems. (発表予定). (2002)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 大脇哲洋: "甲状腺手術における超音波凝固切開装置の問題点とその対策"医科器械学. 70巻10号. 449-450 (2000)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 大脇哲洋: "手術室内の粉塵の分布と減少対策"医科器械学. 71巻. 105-106 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Tetsuhiro Owaki: "The Ultrasonic Coagulating and Cutting System Injures Nerve Function"Endoscopy. (In press). (2002)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Wakako,Rina: "Development of Ultrasonically Activated Bending Scalpel for Endoscopic Surgery - Vibration Analysis in the Bending"Conference on Robotics Mechatronics, Japan Society of Mechanical Engineering. 2A1-D11. (1)-(2) (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Wakako,Rina: "Development of Ultrasonically Activated Bending Scalpel for Endoscopic Surgery - The Basic Concept and the Evaluation of Vibrational State -"19th Annual Conference of Robotics Society of Japan. 955-956 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Wakako,Rina: "Development of Ultrasonically Activated Bending Scalpel for Endoscopic Surgery"Conference on Robotics Mechatronics, Japan Society of Mechanical Engineering. (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Wakako,Rina: "Development of Ultrasonically Activated Bending Scalpel for Endoscopic Surgery"2002 IEEE/RSJ International Conference on Intelligent Robots and Systems. (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Owaki,Tetsuhiro: "The Ultrasonic Coagulating and Cutting System Injures Nerve Function"Endoscopy. in press. (2002)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 