2001 Fiscal Year Final Research Report Summary

OPTIMIZATION OF A CENTRIFUGAL BLOOD PUMP FOR IMPLANTED USE

Research Project

Project/Area Number 12558117
Research Category

Grant-in-Aid for Scientific Research (B)

Allocation TypeSingle-year Grants
Section展開研究
Research Field Biomedical engineering/Biological material science
Research InstitutionRESEARCH INSTITUTE OF NATIONAL CARDIOVASCULAR CENTER

Principal Investigator

TSUKIYA Tomonori  Research Institute of National Cardiovascular Center, Department of Artificial Organs, Research Fellow, 人工臓器部, 室長 (00311449)

Co-Investigator(Kenkyū-buntansha) NISHINAKA Tomohiro  Research Institute of National Cardiovascular Center, Department of Artificial Organs, Division Head, 人工臓器部, 室長 (00256570)
HOMMMA Akihiko  Research Institute of National Cardiovascular Center, Department of Artificial Organs, Research Fellow, 人工臓器部, 室員 (20287428)
TAENAKA Yoshiyuki  Research Institute of National Cardiovascular Center, Department of Artificial Organs, Director, 人工臓器部, 部長 (00142183)
TAKANO Hisateru  Research Institute of National Cardiovascular Center, Vice Dean, 副所長 (60028595)
TATSUMI Eisuke  Research Institute of National Cardiovascular Center, Department of Artificial Organs, Division Head, 人工臓器部, 室長 (00216996)
Project Period (FY) 2000 – 2001
KeywordsVentricular Assist Device / Centrifugal Pump / Hydrodynamic Performance / Hemolysis / Antithrombogenicity / DC Brushless Motor
Research Abstract

The purpose of this study was to optimize the design of the components of the ultracompact centrifugal blood pump which is implanted into the patient as the ventricular assist device. The pump was drastically miniaturized to the half of the former prototype by employing direct drive DC brushless motor.
The impeller design was improved considering the hydrodynamic performances, electric efficiency, washing out condition on the blood-contacting surfaces, and blood cell damage impaired by the pump. The pump performance in vitro satisfied the need for left ventricular assist device. The curved configuration of the inlet duct was also evaluated for better anatomical fitting. The collective performance as a left ventricular assist device was evaluated using an adult goat in chronic animal experiments. 65 days of survival was realized using the pump installed outside the body.
The signals of the driving motor was successfully utilized to estimate the flow rate of the pump with sufficient response. This method can possibly eliminate the use offlowmeter, and also be used to detect the adverse conditions, such as the sucking by excessive pump operation.

  • Research Products

    (6 results)

All Other

All Publications (6 results)

  • [Publications] Tsukiya T, Taenaka Y, Tatsumi E, Takano H.et al.: "Development of an Implantable Centrifugal Blood Pump"Journal of Congestive Heart Failure and Circulatory Support. Vol.1, No.4. 305-308 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Tsukiya T, Taenaka Y, Tatsumi E, Takano H: "Performance of a Newly Developed Implantable Centrifugal Blood Pump"ASAIO Journal. 47. 559-562 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Tsukiya T, Taenaka Y, Tatsumi E, Takano H.et al.: "Application of Indirect Flow Rate Meas urement Using Motor Driving Signals to a Centrifugal Blood Pump with an Integrated Motor"Artificial Organs. Vol.25、No.9. 692-696 (2001)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Tsukiya T, Taenaka Y, Tatsumi E, Takano H: "Performance of a Newly Developed Implantable Centrifugal Blood Pump"ASAIO Journal. 47. 559-562 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] H, Sawairi T, Konishi Y, Masuzawa T, Okada Y: "Development of an Implantable Centrifugal Blood Pump"Journal of Congestive Heart Failure and Circulatory Support. Vol.1,No.4. 305-308 (2001)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Tsukiya T, Taenaka Y, Nishinaka T, Oshikawa M, Ohnishi H, Tatsumi E, Takano H, Konishi Y, Ito K, Shimada M: "Application of Indirect Flow Rate Measurement Using Motor Driving Signals to a Centrifugal Blood Pump with an Integrated Motor"Artificial Organs. Vol.25,No.9. 692-696 (2001)

    • Description
      「研究成果報告書概要(欧文)」より