Three-dimensional Analysis of Cerebral Microvessels and Nerve Cells with Optical Coherence Tomography

Research Project

Project/Area Number 13671479
Research Category

Grant-in-Aid for Scientific Research (C)

Allocation TypeSingle-year Grants
Section一般
Research Field Cerebral neurosurgery
Research InstitutionNational Cardiovascular Center Research Institute

Principal Investigator

SEKI Junji  Natl Cardiovasc Ctr Res Inst, Dept Biomed Eng, Lab Head, 生体工学部, 室長 (20163082)

Co-Investigator(Kenkyū-buntansha) TAKAMIZAWA Keiichi  Natl Cardiovasc Ctr Res Inst, Dept Biomed Eng, Res Assoc, 生体工学部, 室員 (10163312)
NAKAYAMA Yasuhide  Natl Cardiovasc Ctr Res Inst, Dept Biomed Eng, Lab Head, 生体工学部, 室長 (50250262)
Project Period (FY) 2001 – 2003
Keywordsoptical coherence tomography / cerebral cortex / microcirculation / neurovascular coupling / blood flow regulation / Doppler shift / functional column structure / 3-dimensional configuration
Research Abstract

This research project was aimed to analyze 3-dimensional configuration of microvessels and nervous tissue in the cerebral cortex using optical coherence tomography (OCT) system, which is especially designed for in vivo imaging of biological tissues in our laboratory, and to clarify temporal and spatial coupling between neuronal activity and microcirculatory function.
In the OCT system, superluminescent diodes (SLD) with central wavelength of 813 and 940 nm were used for low-coherence light source. The spatial resolutions of the system was evaluated by in vitro and in vivo measurements to be 16x14 and 8x17 μm^2 (longitudinal (depth) x lateral directions) for 813 and 940 nm, respectively. The penetration depth of the OCT system was 0.7-1 mm for the rat cerebral cortex.
The OCT observation revealed that dura and without dura. The responses of the somatosensory cortices during electrical stimulation of the hind paw were also different among the preparation methods ; Delayed swelling of the cortical surface appeared in the somatosensory cortex following the stimulation in the case of dura removal, which was restricted to a thin surface layer with less than several 10 μm. It is considered to reflect the reactive hyperemia accompanying the neuronal activation.
Analyzing Doppler frequency shift due to the blood flow in microvessels by the use of a short time Fourier transform (STFT), cross-sectional velocity profiles and their temporal changes were measured for the rat pial microvessels. While the central velocity in the pial arterioles varied largely with time following the heart beats, the velocity profiles nearly exhibited Poiseuille distribution. These results are consistent with theoretical prediction for the blood flow in microvessels with small Reynold's number and small Womersley parameter. The results of this study indicate that the OCT may provide 3-dimensional information of blood flow distribution in the cerebral cortex with high temporal resolution

  • Research Products

    (18 results)

All Other

All Publications (18 results)

  • [Publications] Satomura Y, Seki J, Ooi Y, Yanagida T, Seiyama A: "In vivo imaging of the rat cerebral microvessels with optical coherence tomography"Clin Hemorheol Microcirc. in press.

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Seki J, Satomura Y, Ooi Y: "Velocity pulse advances pressure pulse by close to 45 degrees in the rat pial arterioles"Biorheology. 41(1). 45-52 (2004)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] 関淳二, 里村泰光: "光コヒーレンス断層法(OCT)による微小循環計測"日本バイオレオロジー学会誌. 17(2). 67-74 (2003)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Seiyama A, Seki J, et al.: "Circulatory basis of fMRI signals : Relationship between changes in the hemodynamic parameters and BOLD signal intensity"NeuroImage. 21(4). 1204-1214 (2004)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Satomura Y, Seki J, Ooi Y, Seiyama A, Yanagida T: "In vivo imaging of the rat brain microvasculature and nervous tissue with optical coherence tomography"Microcirculation Annual. 19. 105-106 (2003)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Satomura Y, Seki J, Ooi Y, Seiyama A, Yanagida T: "In vivo imaging of the rat brain microvasculature with optical coherence tomography"Microcirculation Annual. 18. 127-128 (2002)

    • Description
      「研究成果報告書概要(和文)」より
  • [Publications] Satomura Y, Seki J, Ooi Y, Yanagida T, Seiyama A: "In vivo imaging of the rat cerebral microvessels with optical coherence tomography"Clin Hemorheol Microcirc. (in press).

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Seki J, Satomura Y, Ooi Y: "Velocity pulse advances pressure pulse by close to 45 degrees in the rat pial arterioles"Biorheology. 41(1). 45-52 (2004)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Seki J, Satomura Y: "Microcirculatory measurements with optical coherence tomography (in Japanese)"J Jap Soc Biorheology. 17(2). 67-74 (2003)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Seiyama A, Seki J, Tanabe HC, Sase I, Takatsuki A, Miyauchi S, Eda H, Hayashi S, Imaruoka T, Iwakura T, Yanagida T: "Circulatory basis of fMRI signals : Relationship between changes in the hemodynamic parameters and BOLD signal intensity"NeuroImage. 21(4). 1204-1214 (2004)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Satomura Y, Seki J, Ooi Y, Seiyama A, Yanagida T: "In vivo imaging of the rat brain microvasculature and nervous tissue with optical coherence tomography"Microcirculation Annual. 19. 105-106 (2003)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Seki J, Satomura Y, Ooi Y, Seiyama A, Yanagida T: "Study of brain microcirculation in the rat using optical coherence tomography (in Japanese)"Trans Jpn Soc Med Biol Eng. 41 (Suppl). OS04-01 (2003)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Satomura Y, Seki J, Ooi Y, Seiyama A, Yanagida T: "In vivo imaging of the rat brain microvasculature with optical coherence tomography"Microcirculation Annual. 18. 127-128 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Seki J, Satomura Y, Ooi Y: "Simultaneous measurement of red cell velocity and pressure in the rat pial microvessels"Microcirculation Annual. 18. 3-4 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Taka T, Ohta Y, Seki J, Giddings JC, Yamamoto J: "Impaired flow-mediated vasodilation in vivo and reduced shear-induced platelet reactivity in vitro in response to nitric oxide in prothrombotic, stroke-prone spontaneously hypertensive rats"Pathophysiol Haemost Thromb. 32(4). 184-189 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Ono H, Sasaki Y, Bamba E, Seki J, Giddings JC, Yamamoto J: "Cerebral thrombosis and microcirculation of the rat during the oestrous cycle and after ovariectomy"Clin Exp Pharmacol Physiol.. 29(1-2). 73-78 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Seki J, Satomura Y, Ooi Y, Seiyama A, Yanagida T: "Optical coherence tomography imaging of microvascular and neural networks in the brain"Trans Jpn Soc Med Biol Eng. 40 (Suppl). 33 (2002)

    • Description
      「研究成果報告書概要(欧文)」より
  • [Publications] Satomura Y, Seki J, Seiyama A, Yanagida T: "Optical coherence tomography system developed for the study of microcirculation"Microcirculation Annual. 17. 157-158 (2001)

    • Description
      「研究成果報告書概要(欧文)」より

URL: 

Published: 2005-04-19  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi