• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2007 Fiscal Year Final Research Report Summary

Evaluation of Signal and Noise and Exposure in Medical Image for CAD

Research Project

Project/Area Number 15070205
Research Category

Grant-in-Aid for Scientific Research on Priority Areas

Allocation TypeSingle-year Grants
Review Section Science and Engineering
Research InstitutionNagoya University

Principal Investigator

IKEDA Mitsuru  Nagoya University, Faculty of Medicine, Associate Professor (50184437)

Co-Investigator(Kenkyū-buntansha) ITOH Shigeki  Nagoya University, Faculty of Medicine, Professor (20184658)
IMAI Kuniharu  Nagoya University, Faculty of Medicine, Assistant Professor (20335053)
WADA Shinichi  Niigata University, Faculty of Medicine, Professor (80105519)
NISHIZAWA Kanae  Nagoya University, National Institute of Radiological Sciences, Section Head (40086575)
FURUKAWA Akira  Nagoya University, National Institute of Radiological Sciences, Senior Staff Scientist (70238666)
Project Period (FY) 2003 – 2006
KeywordsImage Quality Evaluation / Signal Detection Theory / Medical Exposure
Research Abstract

We have confirmed that the spatial resolution difference of CT images among CT scanners certainly influenced the diagnostic performance of CAD. To obtain the robustness of CAD for the dependency on CT scanners and improve the performance of CAD, we have developed the method to transform the acquired CT images to the standard ones by using the 3-D LSF which was measured for the considering CT scanner ; and, we have shown that the diagnostic performance of some CADs was improved by this method.
Further, we have assessed the method of estimating image noise SD on CT images, and have shown that the method for estimating the noise variance of an image proposed by Rank et al. will be able to estimate the CT image noise SD with reasonable accuracy, unless the neighborhood pixels does not too much correlate with each other. We have also shown that streak artifacts on CT images can be statistically modeled by the Gumbel distribution, and, by using this result, we have devised the new evaluation method of streak artifacts on CT images that we call as the Gumbel evaluation method. Further more, we have assessed whether or not the fractal-feature distance could be a substitute for the observer performance index in Contrast-Detail (C-D) examination and image-reading study for evaluation of anatomic noise ; and, we have shown that all C-D and chest images can be analyzed by the virtual volume method and that the fractal-feature distances would be equivalent to IQFs and observer's confidence levels.
In addition, we have measured the dose distributions of various organs and tissues during examinations with multi-detector row CT, under the irradiation conditions close to actual clinical examinations, and derived the corresponding effective doses. Further, we have evaluated the relationship between radiation dose and various image quality indices.

  • Research Products

    (9 results)

All 2007 2006 2005

All Journal Article (7 results) (of which Peer Reviewed: 3 results) Presentation (1 results) Book (1 results)

  • [Journal Article] Fractal-feature distance analysis of radiographic image2007

    • Author(s)
      Kuniharu Imai, et al.
    • Journal Title

      Academic Radiology 14-2

      Pages: 137-143

    • Description
      「研究成果報告書概要(和文)」より
    • Peer Reviewed
  • [Journal Article] Fractal-feature distance analysis of radiographic image2007

    • Author(s)
      Kuniharu Imai, et al
    • Journal Title

      Academic Radiology 14-2

      Pages: 137-143

    • Description
      「研究成果報告書概要(欧文)」より
  • [Journal Article] An effective method to verify line and point spread functions measured in computed tomography2006

    • Author(s)
      Masaki Ohkubo, et al.
    • Journal Title

      Medical Physics 33-8

      Pages: 2757-2764

    • Description
      「研究成果報告書概要(和文)」より
    • Peer Reviewed
  • [Journal Article] An effective method to verify line and point spread functions measured in computed tomography2006

    • Author(s)
      Masaki Ohkubo, et al
    • Journal Title

      Medical Physics 33-8

      Pages: 2757-2764

    • Description
      「研究成果報告書概要(欧文)」より
  • [Journal Article] Enlarged longitudinal dose profiles in cone・beam CT and the need for modified dosimetry2005

    • Author(s)
      Shinichiro Mori, et al.
    • Journal Title

      Medical Physics 32-4

      Pages: 1061-1069

    • Description
      「研究成果報告書概要(和文)」より
    • Peer Reviewed
  • [Journal Article] Enlarged longitudinal dose profiles in cone-beam CT and the need for modified dosimetry2005

    • Author(s)
      Shinichiro Mori, et al
    • Journal Title

      Medical Physics 32-4

      Pages: 1061-1069

    • Description
      「研究成果報告書概要(欧文)」より
  • [Journal Article] Chapter 4 Radiation Exposure and Risk Assessment in HRCT : International classification of HRCT for occupational and environmental respiratory disease2005

    • Author(s)
      Kanae Nishizawa, et al
    • Journal Title

      Springer, Tokyo

    • Description
      「研究成果報告書概要(欧文)」より
  • [Presentation] The CT image standardization based on the verified PSF2007

    • Author(s)
      Shinichi Wada, et al.
    • Organizer
      SPIE Medical Imaging 2007
    • Place of Presentation
      San Diego, USA
    • Year and Date
      20070217-20070222
    • Description
      「研究成果報告書概要(和文)」より
  • [Book] Chapter 4 Radiation Exposure and Risk Assessment in HRCT : International classification of HRCT for occupational and environmental respiratory disease2005

    • Author(s)
      Kanae Nishizawa, et al.
    • Total Pages
      145
    • Publisher
      Springer, Tokyo
    • Description
      「研究成果報告書概要(和文)」より

URL: 

Published: 2010-06-09  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI