2017 Fiscal Year Final Research Report
Development of integrated system of wearable EEG device and head movement and EOG measurement device
Project/Area Number |
15H02743
|
Research Category |
Grant-in-Aid for Scientific Research (B)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Human interface and interaction
|
Research Institution | National Institute of Information and Communications Technology |
Principal Investigator |
Naruse Yasushi 国立研究開発法人情報通信研究機構, 脳情報通信融合研究センター脳情報工学研究室, 室長 (00455453)
|
Co-Investigator(Kenkyū-buntansha) |
横田 悠右 国立研究開発法人情報通信研究機構, 脳情報通信融合研究センター脳情報工学研究室, 研究員 (10710593)
|
Project Period (FY) |
2015-04-01 – 2018-03-31
|
Keywords | ウェアラブル脳波計 / GPS |
Outline of Final Research Achievements |
We succeeded in developing a prototype electroencephalogram (EEG) device with a time synchronization system using a GPS unit. Since the GPS unit receives accurate time information from the satellite, it can supply a synchronization signal in the order of microseconds or more. By using the GPS unit, it is possible to synchronize on the order of microseconds even via a wireless connection. By using the multiple prototype EEG devices with the time synchronization system using the GPS unit, it became possible to synchronously measure EEG and electro-oculogram of multiple people via wireless connection. Then, we measured brain activities of players playing match-up video games. As a result, we succeeded in measuring the brain activity that changes according to the situation of the game.
|
Free Research Field |
脳機能計測
|