2017 Fiscal Year Final Research Report
Safety evaluation of large-size curved balloon catheter for clinical use
Project/Area Number |
15K19608
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Multi-year Fund |
Research Field |
Pediatrics
|
Research Institution | University of Yamanashi |
Principal Investigator |
KISE Hiroaki 山梨大学, 大学院総合研究部, 助教 (40436879)
|
Research Collaborator |
TOMITA Hideshi
FUJII Takanari
FUJIMOTO Kazuto
KONO Yosuke
|
Project Period (FY) |
2015-04-01 – 2018-03-31
|
Keywords | 小児循環器 / カテーテル治療 / ステント留置 / 屈曲病変 / 狭窄病変 / curved balloon / 後拡張 / 術後狭窄 |
Outline of Final Research Achievements |
We demonstrated the dilatability, the sufficient angle of the 8mm-curved balloon catheter. The 8 mm-curved balloon catheter was approved to be deflated in vivo and safely removed through a introducer sheath. During balloon dilation, the curved balloon catheter was fixed at the stenotic site without slipping, while straight balloon catheter slipped easily. In stenting for curved lesion using a straight balloon catheter, vessel deformity and stent mal-apposition were appeared on the greater and lesser curvatures, respectively, which were completely resolved by post-dilatation using curved balloon catheter.
|
Free Research Field |
小児循環器領域
|