Outline of Final Research Achievements |
Patient's respiratory motion will be changed due to the baseline drift. The accuracy of a prediction model for the dynamic tumor tracking will be decreased as the treatment time elapsed. We have developed four-axis moving phantom which enables to reproduce the patient’s respiratory motion including the baseline drift with high accuracy. The novel prediction model for the dynamic tumor tracking improved the prediction accuracy by updating the model using the real tumor position during beam delivery. However, the sampling interval greater than 2 seconds have a risk for creating the low-accurate prediction model.
|