2017 Fiscal Year Final Research Report
Establishment of core technology for low-dose and multi-color X-ray CT system
Project/Area Number |
16H07266
|
Research Category |
Grant-in-Aid for Research Activity Start-up
|
Allocation Type | Single-year Grants |
Research Field |
Medical systems
|
Research Institution | Kanazawa University (2017) Waseda University (2016) |
Principal Investigator |
|
Project Period (FY) |
2016-08-26 – 2018-03-31
|
Keywords | X線CT / MPPC / LSI / 多色 / 低被ばく |
Outline of Final Research Achievements |
X-ray CT is a basic technology to support the modern healthcare, but a large radiation dose by X-ray CT is a serious issue. To overcome the difficulty, we newly developed an X-ray CT system using multi-pixel photon counters (MPPCs), instead of photodiodes which have been used for conventional X-ray CT. We found that the MPPC CT system can drastically reduce the radiation dose by a factor of ~100 compared with the conventional X-ray CT, thereby indicating great potential for low-dose CT imaging. Furthermore, we demonstrated the multi-color X-ray CT imaging and successfully developed a multi-channel MPPC array system. These results show that we succeeded in establishment of the core technology for future low-dose and multi-color X-ray CT using MPPCs.
|
Free Research Field |
放射線検出器
|