• Search Research Projects
  • Search Researchers
  • How to Use
  1. Back to project page

2017 Fiscal Year Final Research Report

Development of a low-price direction sensitive radiation detector

Research Project

  • PDF
Project/Area Number 16K13796
Research Category

Grant-in-Aid for Challenging Exploratory Research

Allocation TypeMulti-year Fund
Research Field Particle/Nuclear/Cosmic ray/Astro physics
Research InstitutionYamagata University

Principal Investigator

Nakamori Takeshi  山形大学, 理学部, 准教授 (30531876)

Research Collaborator GUNJI Shuichi  山形大学, 理学部, 教授 (70241685)
Project Period (FY) 2016-04-01 – 2018-03-31
KeywordsMPPC / ガンマ線 / 方向検知 / Raspberry Pi / Android
Outline of Final Research Achievements

We develop a small, light-weigh and low-price gamma ray detector which is sensible to source directions. The detector consists of a 3-cm cubic lead block with its all faces having GAGG:Ce scintillators, and calculate the source directions from a combination of count rates of the scintilators. We established the designed accuracy of 22.5 degree. We also develop a data acquisition system and analysis softwares on a card-type PC, Raspberry Pi, and performed a demonstration outside. For further convenience, we apply a 4-channel ADC board and optimize a detector structure. Finally we develop an application running on a smart phone to operate and monitor the whole system.

Free Research Field

放射線物理学

URL: 

Published: 2019-03-29  

Information User Guide FAQ News Terms of Use Attribution of KAKENHI

Powered by NII kakenhi