2019 Fiscal Year Final Research Report
Physical study of simple and highly conformal total body irradiation
Project/Area Number |
17K10499
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Multi-year Fund |
Section | 一般 |
Research Field |
Radiation science
|
Research Institution | National Cancer Center Japan |
Principal Investigator |
Naoki Nakamura 国立研究開発法人国立がん研究センター, 東病院, 医長 (40359625)
|
Co-Investigator(Kenkyū-buntansha) |
橘 英伸 国立研究開発法人国立がん研究センター, 先端医療開発センター, 医学物理専門職 (20450215)
|
Project Period (FY) |
2017-04-01 – 2020-03-31
|
Keywords | 全身照射 / 強度変調放射線治療 / 白血病 / 骨髄移植 / パッチ照射 |
Outline of Final Research Achievements |
Total body irradiation (TBI) is commonly used as a induction therapy of bone marrow transportation for leukemia. The field length of linear accelerator is at most 40 cm. Therefore, inordinately method of radiotherapy is warranted for TBI. Many institutions use long SAD technique or moving couch technique. However, such techniques have shortcomings in accuracy, safety, consuming time, or cost. To improve such problems, we developed patchy VMAT technique for TBI.
|
Free Research Field |
放射線治療
|
Academic Significance and Societal Importance of the Research Achievements |
我々はパッチVMAT法が時間的、技術的に臨床導入可能な照射方法であることを検証した。今後臨床試験を経て、臨床導入されれば、全身やリスク臓器の線量がより正確に計算可能となり、出力値や照射野形状を最適化することにより良好な線量分布を作成することができる、かつ通常照射と同様の照射方式を用いるためエラーが生じるリスクも低く安全で準備時間短縮も可能である。また、パッチVMAT法は通常の放射線治療設備で実施可能であり高価な専用の治療寝台を購入する必要もなく低コストである。
|