2007 Fiscal Year Final Research Report Summary
Development of real-time image guided needle insertion robot using MRI and Ultrasound
Project/Area Number |
18500380
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Medical systems
|
Research Institution | Kyushu University |
Principal Investigator |
HONG Jaesung Kyushu University, Faculty of Medical Sciences, Visiting Assistant Professor (70404043)
|
Co-Investigator(Kenkyū-buntansha) |
MATUMIYA Kiyoshi Tokyo University, Graduate school of information Science and Technology, Assistant Professor (40396792)
KONISHI Kouzou Kyushu University, Faculty of Medical Sciences, Visiting Assistant Professor (90380641)
NAKASHIMA Hideaki Kyushu University, Faculty of Medical Sciences, Visiting Researcher (20253528)
HASHIZUME Makoto Kyushu University, Faculty of Medical Sciences, Professor (90198664)
|
Project Period (FY) |
2006 – 2007
|
Keywords | Surgical Navigation / Needle Insertion Robot / Image-guided surgery / レジストレーション |
Research Abstract |
1. Integration of Ultrasound and MRI Using optical position sensor, identical cross-sectional MR images were displayed with real-time ultrasound images, so that surgeons can do the therapies more accurately and safely. The position sensor makers were attached to the ultrasound probe. The needle position and orientation were calculated from the constant positional relationship between the probe and the needle. The needle tip position that is difficult to identify in ultrasound images, could be estimated by the developed system. 2. Hybrid auto-registration Since skin affixed markers were influenced by skin deformation, anatomical landmarks were used at the same time. This hybrid method significantly produced high registration accuracy compared to the single use. We also developed an automatic matching algorithm between the markers in image and patient spaces. The doctors don't have to remember the order of markers, so that the registration can be completed fast and precisely. The color interface also presents the error of each marker; therefore user can select and exclude the specific marker. Users can also select the automatic or manual procedure for the registration. 3. Prototype of insertion robot A prototype of insertion robot was developed for the purpose of automatic needle direction decision from the coordinate of navigation system. This robot has 2 degrees of freedom to take orientation in 3-D space. This robot does not have xyz stage, so that it has small size and light weight. Instead of xyz position determination, human interference is involved. This robot has the concept of using 2-D ultrasound image and 3-D MR images. It has two micro motors inside body; therefore, portable use is also possible. 4. Clinical applications The navigation system was used for radio frequency ablations and ethanol injection therapies for about 40 tumor nodules. It was especially useful for target identification, when the ultrasound images could not show the target tumor clearly.
|
-
-
-
-
-
[Journal Article] An NET Navigation System Displaying Intraoperative Virtual CT.2007
Author(s)
Hong, J., Matsumoto, N., Ouchida, R., Komune, S., Hashizume, M
-
Journal Title
The Institute of Electronic, Information and Communication Engineers Vol. 107(220)
Pages: 17-21
Description
「研究成果報告書概要(欧文)」より
-
-
-
[Journal Article] Interventional navigation for abdominal surgery by simultaneous use of MRI and ultrasound2006
Author(s)
Hong, J., Nakashima, H., Konishi, K., Ieiri, S., Tanoue, K., Hashizume, M
-
Journal Title
Med Bio Eng Comput Vol. 44(12)
Pages: 1127-1134
Description
「研究成果報告書概要(欧文)」より
-
-
-
-
-
-
-
-
-
[Presentation] Multi-modality Surgical Navigation for Percutaneous Ethanal Injection Therapy for HCC.2006
Author(s)
Hong, J., Nakashima, H., Kotou, K., Katou, M., Enjyouji, S., Hashizume, M
Organizer
Japan Society of Computer Aided Surgery
Year and Date
20060000
Description
「研究成果報告書概要(欧文)」より
-
[Presentation] Image-guided abdominal surgery by integration of MRI and ultrasound.2006
Author(s)
Hong, J., Konishi, K., Nakashima, H., Ieiri, S., Tanoue, K., Hashizume, M
Organizer
IFMBE Proceedings
Year and Date
20060000
Description
「研究成果報告書概要(欧文)」より
-
[Presentation] Integration of MRI and ultrasound in surgical navigation for robotic surgery.2006
Author(s)
Hong, J., Konishi, K., Nakashima, H., Ieiri, S., Tanoue, K., Nakamuta, M., Hashizume, M
Organizer
IFMBE Proceedings
Year and Date
20060000
Description
「研究成果報告書概要(欧文)」より
-
[Presentation] Surgical navigation for abdominal surgery based on integration of MRI with ultrasound.2006
Author(s)
Hong, J., Konishi, K., Nakashima, H., Ieiri, S., Tanoue, K., Hashizume, M
Organizer
Int J CARS
Year and Date
20060000
Description
「研究成果報告書概要(欧文)」より
-
[Presentation] Developmentandclnicalapplicationofsurgical navigation system for abdominal approach using MRI and ultrasound.2006
Author(s)
Hong, J., Nakashima, H., Konishi, K., Kotou, K., Katou, M., Enjyouji, S., Hashizume, M
Organizer
The 6th Japan Neurosugery, Research Meeting for Intraoperative Image
Year and Date
20060000
Description
「研究成果報告書概要(欧文)」より
-
[Presentation] Image guided therapeutic system based on integration of Open MRI and ultrasound.2006
Author(s)
Hong, J., Nakashima, H., Kotou, K., Katou, M., Enjyouji, S., Hashizume, M
Organizer
The 1st Research Meeting for Liver Simulation
Year and Date
20060000
Description
「研究成果報告書概要(欧文)」より
-
-
-
-
-
-