2021 Fiscal Year Final Research Report

Development of real time monitoring system of radiotherapy accuracy by Cherenkov emission

Research Project

  • PDF
Project/Area Number 18K15613
Research Category

Grant-in-Aid for Early-Career Scientists

Allocation TypeMulti-year Fund
Review Section Basic Section 52040:Radiological sciences-related
Research InstitutionNational Cancer Center Japan

Principal Investigator

Okamoto Hiroyuki  国立研究開発法人国立がん研究センター, 中央病院, 医学物理士 (90595729)

Project Period (FY) 2018-04-01 – 2022-03-31
Keywords放射線治療 / 品質管理 / チェレンコフ光 / フィルム
Outline of Final Research Achievements

In this study, we developed a real tie monitoring system to evaluate radiotherapy accuracy using Cherenkov emission generated by irradiation. The Cherenkov emission detector is a C-dose camera (OptiDose, Inc.), which equips a scintillator built into the camera to detect scattered radiation and images with a high signal-to-noise ratio as a trigger signal can be obtained. By using this method, Cherenkov emission could be observed even in relatively bright treatment rooms. Comparison with conventional radiation detector, radiochromic film was also performed, and the intensity distribution was almost similar to that of the film. The linearity with the irradiation MU was confirmed, and it was demonstrated that it is possible to perform dosimetric assessment.

Free Research Field

放射線物理

Academic Significance and Societal Importance of the Research Achievements

従来では放射線照射中の体表面の線量を計測する場合には皮膚表面に放射線に感度を持つフィルムや熱ルミネッセンス線量計,ガラス線量計などを貼付する必要があるが,照射中のチェレンコフ光を観測することにより非侵襲的に表面線量を評価することが可能となった.この方法が確立すれば,患者の負担,線量計読み出しのための後処理作業を減らすことができる.品質管理ツールにおいては,従来フィルムで照射位置を評価していたが,チェレンコフ光を用いることで照射位置のみならず線量情報,放射線のエネルギーも評価できる.またすべて画像処理で結果を解析するため,フィルムなどの消耗品も減らすことができる.