2010 Fiscal Year Final Research Report
Development of a new dental unit without a water supply pipe for infection control.
Project/Area Number |
19390489
|
Research Category |
Grant-in-Aid for Scientific Research (B)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
補綴理工系歯学
|
Research Institution | Tohoku University |
Principal Investigator |
TAMAZAWA Yoshinori Tohoku University, 病院, 准教授 (10124603)
|
Co-Investigator(Kenkyū-buntansha) |
TAMAZAWA Kaoru 東北大学, 病院, 講師 (00124602)
IWAMATSU Masaaki 東北大学, 病院, 助教 (30343031)
WATANABE Makoto 東北大学, 大学院・歯学研究科, 客員教授 (80091768)
|
Project Period (FY) |
2007 – 2010
|
Keywords | 感染対策 / 歯科用ユニット / 給水管 / バイオフィルム / 細菌 / 着脱式タンク |
Research Abstract |
Water contamination from dental unit is a serious problem. So, we developed a dental unit installing a built-in removable water supply tank, without a water supply pipe to eliminate biofilm, the source of water contamination in the pipe. And, we grew common bacteria on standard agar medium at 36℃ for one days, and heterotrophic bacteria on R2A agar medium at 20℃ for 7 days, and measured bacterial numbers for water quality tests based on the conventional method. The results showed that the number of common bacteria from turbine, engine-handpiece and three-way syringe ranged between 30-50 CFU/ml and heterotrophic bacteria ranged between 30-1,500 CFU/ml, achieving the standard value of water quality ( common bacteria : 100 CFU/ml, heterotrophic bacteria : 2,000 CFU/ml) in Japan.
|