2009 Fiscal Year Final Research Report
Nanostructure surface modification of the orthodontic appliance by a plasma immersion ion implantation
Project/Area Number |
19592363
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Orthodontic/Pediatric dentistry
|
Research Institution | Health Sciences University of Hokkaido |
Principal Investigator |
IIJIMA Masahiro Health Sciences University of Hokkaido, 歯学部, 准教授 (20305915)
|
Co-Investigator(Kenkyū-buntansha) |
MIZOGUCHI Itaru 北海道医療大学, 歯学部, 教授 (20200032)
|
Research Collaborator |
BRANTLEY William A. Ohio State University
LIN Chun-pin National Taiwan University
|
Project Period (FY) |
2007 – 2009
|
Keywords | イオン注入 / 矯正装置 / 耐食性 |
Research Abstract |
The purposes of this study were to investigate the corrosion properties and the surface structure of ion implanted nickel-titanium orthodontic wire. Ion implanted nickel-titanium wire and Non ion implanted nickel-titanium wire were used for a potentiodynamic polarization measurement and X-ray photoelectron spectroscopy (XPS). Ion implanted wire in artificial saliva and mouthrinse with neutral pH had a lower current density than non ion implanted wire. Both wires in mouthrinse with low pH showed similar corrosion properties. XPS results suggested that a TiN layer may be present on the top surface. Plasma immersion ion implantation might increase the corrosion resistance of nickel-titanium wire in a neutral environment because of a protective TiN layer on the top surface.
|