2020 Fiscal Year Final Research Report
Development of pregnant / fetal model (phantom) and radiation dose reduction
Project/Area Number |
19K17223
|
Research Category |
Grant-in-Aid for Early-Career Scientists
|
Allocation Type | Multi-year Fund |
Review Section |
Basic Section 52040:Radiological sciences-related
|
Research Institution | Tohoku University |
Principal Investigator |
Matsunaga Yuta 東北大学, 医学系研究科, 大学院非常勤講師 (30803019)
|
Project Period (FY) |
2019-04-01 – 2021-03-31
|
Keywords | 胎児線量 / 線量低減 / コンピュータ断層撮影 / 被曝 |
Outline of Final Research Achievements |
The phantom developed in this study has two types of abdominal circumference sizes, and is designed so that a dosimeter can be placed in the phantom, and we have developed an unprecedented pregnant woman's abdominal phantom. Furthermore, using the developed phantom, we evaluated the radiation dose on prenatal diagnosis of skeletal dysplasias with CT and examined measures to reduce the fetal radiation dose. Regarding measures to reduce fetal radiation dose, it was confirmed that the radiation dose was reduced compared to when the voltage was low by setting the tube voltage at the time of imaging high.
|
Free Research Field |
放射線技術学
|
Academic Significance and Societal Importance of the Research Achievements |
妊娠子宮の大きさは、週数や体格によっても異なるため、2種類の腹囲サイズのファントムを開発した。開発したファントムは内部に線量計を配置できるようにし、胎児の線量測定が可能になった。胎児被曝線量の低減策に関して、撮影時の管電圧を高く設定することで低電圧時に比べて被曝線量の低減を確認した。低電圧撮影の利点として、低コントラスト分解能向上やヨード造影剤使用時の造影効果の増強が挙げられる。しかし、骨系統疾患の出生前診断では高コントラストである骨の評価を行い、ヨード造影剤を使った検査は一般的に行われない。これらの結果を踏まえて、CTによる骨系統疾患の出生前診断の管電圧は、高電圧に設定することを提案したい。
|