2009 Fiscal Year Final Research Report
Development of a Light sharing DOI-PET detector using GAPD
Project/Area Number |
20700416
|
Research Category |
Grant-in-Aid for Young Scientists (B)
|
Allocation Type | Single-year Grants |
Research Field |
Medical systems
|
Research Institution | National Institute of Radiological Sciences |
Principal Investigator |
NISHIKIDO Fumihiko National Institute of Radiological Sciences, 分子イメージング研究センター, 研究員 (60367117)
|
Project Period (FY) |
2008 – 2009
|
Keywords | 検査・診断システム / PET装置 |
Research Abstract |
Geiger-mode avalanche photodiodes (APDs) have many advantages to be used in PET detector, such as high internal gain, high photon detection efficiency, lower power supply and insensitivity to magnetic fields. We constructed prototype four-layer DOI detectors using two MPPC arrays and evaluated their detector performances. As a result, better detector performances as a PET detector than a detector using PS-PMT was achieved.
|