2011 Fiscal Year Final Research Report
Development of the less invasive cardiac imaging method using 320-detector row CT
Project/Area Number |
21591576
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Radiation science
|
Research Institution | Iwate Medical University |
Principal Investigator |
|
Co-Investigator(Renkei-kenkyūsha) |
TANAKA Ryoichi 岩手医科大学, 医学部, 講師 (30393218)
|
Project Period (FY) |
2009 – 2011
|
Keywords | 放射線 / 循環器 / 高血圧 / 臨床 / CT |
Research Abstract |
We developed the low radiation dose cardiac imaging method by "single-beat scan with prospective ECG-gating" using a 320-detector row CT. The diagnostic accuracy of this method for coronary in-stent restenosis was 100% of sensitivity, specificity 87.8% in consecutive 102 patients, and the effective radiation dose was approximately 4.8 mSv. We could reduce approximately 32% of radiation dose if we use the adaptive iterative reconstruction method. We tried the subtraction coronary CT using this technique for the patients who had severe coronary artery calcifications.
|