2011 Fiscal Year Final Research Report
Establishing regional dosimetry protocol for panoramic radiography
Project/Area Number |
21592399
|
Research Category |
Grant-in-Aid for Scientific Research (C)
|
Allocation Type | Single-year Grants |
Section | 一般 |
Research Field |
Pathobiological dentistry/Dental radiology
|
Research Institution | Showa University |
Principal Investigator |
SAKAINO Rie 昭和大学, 歯学部, 助教 (50153862)
|
Project Period (FY) |
2009 – 2011
|
Keywords | 歯科放射線学一般 |
Research Abstract |
To establish regional diagnostic reference levels in panoramic radiography and to promote optimization of radiation protection for patients, dose-width products DWPs should be widely surveyed. For measurement of the DWPs in panoramic radiography we improved optically stimulated luminescence(OSL) dosimeters for CT. Using the improved OSL dosimeter, we measured the DWP, and at the same time the beam height H using X-ray films. A dose-area product DAP was calculated from the product of DWP and H. We evaluated the accuracy of the DWP and DAP values comparing with calibrated ionization chamber measurement, and concluded the validity of the new regional dosimetry protocol for panoramic radiography.
|